排序方式: 共有1条查询结果,搜索用时 0 毫秒
1
1.
A multifunctional HTDMA system with a robust temperature control 总被引:3,自引:0,他引:3
The hygroscopicity of atmospheric aerosols significantly influences their size distribution,
cloud condensation nuclei ability, atmospheric residence time, and climate forcing. In order to investigate
the hygroscopic behavior of aerosol particles and serious haze in China, a Hygroscopic Tandem Differential
Mobility Analyzers (HTDMA) system was designed and constructed at Fudan University. It can function as a
scanning mobility particle sizing system to measure particle size distribution in the range of 20--1000 nm
in diameter, as well as a hygroscopicity analyzer for aerosol particles with diameters between 20--400 nm
in the range of 20%--90% RH (relative humidity). It can also measure the effect of uptake of inorganic acids
or semi-VOCs on the hygroscopic behavior of aerosols, such as typical inorganic salts in atmospheric dust or
their mixtures. The performance tests show that the system measured particle size of the standard polystyrene
latex spheres (PSLs) is 197 nm, which is in excellent agreement with the certified diameter D=199±6 nm, as
well as a standard deviation of the repeated runs SD=8.9x10-4. In addition, the measured hygroscopic
growth factors of the model compounds, (NH4)2SO4 and NaNO3, agree with the Kohler theoretical
curves. The results indicate that the HTDMA system is an excellent and powerful tool for studying the
hygroscopic behavior of submicron aerosols and meets the demand required for laboratory research and
fieldwork on atmospheric aerosols in China. 相似文献