SOMATOM DRH CT机US0080故障分析与排除
引用本文:张延武. SOMATOM DRH CT机US0080故障分析与排除[J]. CT理论与应用研究, 1994, 3(2): 40-41
作者姓名:张延武
作者单位:广州军区总医院
摘    要:CT机在正常扫描情况下,图象上出现了一簇斜方向伪影。同时屏显US0080错误信息。从提示的错误信息分析,会有许多原因导致US0080信息出现.本文讨论了由于GCP信号瞬间丢失而产生的图象伪影。以及修复过程同时也找出了,由于BSP内5V电源低于正常值而产生的图象上有点状伪影的故障原因。

关 键 词:伪影  GCP信号  CT  故障分析  排除  医用

Analysis and Elimination of US0080 Error Information in SOMATOM DRH CT Set
Zhang Yanwu. Analysis and Elimination of US0080 Error Information in SOMATOM DRH CT Set[J]. Computerized Tomography Theory and Applications, 1994, 3(2): 40-41
Authors:Zhang Yanwu
Abstract:During CT Scanning there was a normal image in the direction of CCW.However in direction of CW the obliquestrenks appeared and US0080 mistake was showed on the screan.Author found:when the gunery rotated in the direction of CW.GCP single lamp go out for a moment,because GCP single line broke in a twinkling.Aften the line was changed the scanning progress recovered,but there were some white dotted artifacts around the image when SV power in BSP was increased from 4.93 V to 5.03 V,actted artifacts mentioned above disapeared.
Keywords:s:Elimination   Error information