Measurement of trace nitrate concentrations in seawater by ion chromatography with valve switching |
| |
Authors: | Juan Du Yun Fa Yue Zheng Xuebing Li Fanglin Du Haiyan Yang |
| |
Affiliation: | 1. Public Laboratory, Qingdao Institute of Bioenergy and Bioprocess Technology, Chinese Academy of Sciences, Qingdao, 266101, China 2. College of Materials Science and Engineering, Qingdao University of Science & Technology, Qingdao, 266042, China 3. College of Food Science and Engineering, Shandong Agricultural University, Tai’an, 271018, China 4. Heterogeneous Catalysis Conversion Group, Qingdao Institute of Bioenergy and Bioprocess Technology, Chinese Academy of Sciences, Qingdao, 266101, China
|
| |
Abstract: | An ion chromatographic method with a valve switching facility was developed to determine trace nitrate concentrations in seawater using two pumps, two different suppressors, and two columns. A carbohydrate membrane desalter was used to reduce the high concentrations of sodium salts in samples. In this method, trace nitrate was eluted from the concentrator column to the analytical columns, while the matrix fl owed to waste. Neither chemical pre-treatment nor sample dilution was required. In the optimized separation conditions, the method showed good linearity ( R >0.99) in the 0.05 and 50 mg/L concentration range, and satisfactory repeatability (RSD<5%, n =6). The limit of detection for nitrate was 0.02 mg/L. Results showed that the valve switching system was suitable and practical for the determination of trace nitrate in seawater. |
| |
Keywords: | ion chromatography valve switching seawater nitrate ion |
本文献已被 CNKI 维普 SpringerLink 等数据库收录! |
|