Abstract: | A preliminary study was conducted to outline the laser Raman spectroscopy technique for determination of salinity in the aqueous phase in fluid inclusions. The skewing parameters of the Raman profiles of the calibration solutions determined were used to derive a calibration curve for the estimation of the equivalent mass fraction NaCl in aqueous solutions. This technique was also verified in the analysis of the natural fluid inclusions from Tongshankou porphyry Cu (Mo) deposit, Hubei Province, China. Although the analyses on the natural fluid inclusions are limited, an acceptable agreement has been reached on the salinities, for most fluid inclusions, determined by the Raman spectroscopy and microthermometry techniques, indicating the reliability of the Raman technique for determination of salinity in fluid inclusion studies. |