首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Development and assessment of an analytical system for the accurate and continual measurement of total dissolved inorganic carbon
Authors:Carol Robinson  P J leB Williams
Abstract:An automated coulometric titration system based on that described by Johnson, Sieburth, Williams and Brandstrom (1987, Mar. Chem., 21: 117–133) has been evolved for the accurate and continual measurement of total dissolved inorganic carbon (TCO2). The instrument achieves an analytical precision (1 SD) of ±0.5–1.0 μmol kg−1 (0.025–0.05%). The accuracy of the system has been examined by a limited comparison with other coulometric-based titrators and with a manometric-based system; agreement was to 1 μmol kg−1. The capability for automatic continual analysis allows surface mapping of TCO2; a sample rate of 10 analyses h−1 gives a mapping resolution of 1–2 km. Provision for frequent standardization with a liquid substandard has been included in the development. The ability to achieve high-density analyses while maintaining interlaboratory consistency and standardization constitutes a vital contribution to surveys of ocean carbon chemistry (e.g. Joint Global Ocean Flux Study, World Ocean Circulation Experiment).
Keywords:
本文献已被 ScienceDirect 等数据库收录!