首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Chemical analysis of mineral dust grains by SIMS: Application to a cometary mission
Authors:N Pailer  E Grün  D Bahr  D Lang
Institution:Max-Planck-Institut für Kernphysik, Postfach 103980, D-6900 Heidelberg, Federal Republic of Germany
Abstract:Analysis of a dust sample (e.g. collected during a cometary rendezvous mission) by SIMS (Secondary Ion Mass Spectroscopy) can provide information on elemental abundances (? 100 amu), the molecular composition of grain surfaces, and isotopic ratios of selected elements. This can be accomplished with dust covering as little as 10?4 of the collector surface area. In order to demonstrate these capabilities a special experimental set-up for substrate preparation, dust collection and SIMS analysis of dust under ultrahigh vacuum conditions was developed. The comparison of elemental abundance ratios for different olivines and pyroxenes measured with the special SIMS equipment with that measured by an electron microprobe indicated an accuracy for SIMS of the elemental abundance measurements of ? 30%. By varying the energy threshold of secondary ions to be mass-analysed from 0 to 50 eV it is possible to identify molecular ions in the spectra and to estimate their abundance with respect to elemental ions on the same mass line. The ratios of molecular to elemental ions vary by a factor of 1–25. The concept for a future cometary rendezvous experiment as well as first results of chemical investigation on mineral dust samples obtained are reported.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏