Analysis of noble gases in water using a quadrupole mass spectrometer in static mode |
| |
Authors: | Jason C. Poole Gavin W. McNeill Stephen R. Langman Frank Dennis |
| |
Affiliation: | Postgraduate Research Institute for Sedimentology, University of Reading, Whiteknights, Reading, Berkshire RG6 6AB, U.K. |
| |
Abstract: | A quadrupole-based mass spectrometer used in static mode has been employed to determine noble gas concentrations of water samples by isotope dilution. Water samples are degassed and spiked with minor isotopes of each noble gas. After separation of the various gas components, the isotopic ratios are measured in the mass spectrometer. Because a spike is added to each sample, the recovery of each gas component does not have to be quantitative. Equilibration temperatures are calculated from the gas concentrations, based on noble gas solubility data. An accuracy of ±1°C, from the air equilibration temperature, is attainable for air-saturated water samples prepared under controlled conditions in the laboratory. The method has been applied to groundwaters from the London Basin Chalk aquifer. The noble-gas-derived temperatures are consistent with climatic conditions prevailing at the periods of recharge. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|