An open path,fast response infrared absorption gas analyzer for H2O and CO2 |
| |
Authors: | David L. Auble Tilden P. Meyers |
| |
Affiliation: | (1) Atmospheric Turbulence and Diffusion Division, National Oceanic and Atmospheric Administration, Air Resources Laboratory, P.O. Box 2456, 37831 Oakridge, TN, USA |
| |
Abstract: | An open path infrared absorption based instrument for fast response measurements of H2O and CO2 fluctuations is described. This instrument performed reliably in several field experiments in both terrestrial and marine environments, on both fixed (tower) and mobile (boat, plane) flux platforms. Noise levels for H2O and CO2 concentrations were less than 10 mg/m3 and 300 g/m3, respectively for frequencies between 0.005 and 10 Hz. Drifts in instrument output, associated with changes in instrument temperature, are compensated for electronically. |
| |