Determination of Selenium in Geochemical Reference Samples Using Flameless Atomic Absorption Spectrometry
Authors:J. ERZINGER  H. PUCHELT
Affiliation:Institut für Petrographie und Geochemie der Universität Karlsruhe, Kaiserstr. 12, D-7500 Karlsruhe, BRD
Abstract:Sixteen international and six inter-laboratory geochemical reference samples have been analysed for their selenium content. No previous Sedata are available for thirteen of them. Selenium has been separated from silicate matrix by volatilization and was determined with flameless atomic absorption spectrometry. An outline of the procedure is given. The limit of detection is 10 ng Se per 1 g of sample. Time required for one full analysis is 1.2 to 2.5 hours depending on amount of sample to be volatilized.