对修订现行工业CT相关国家标准的建议 |
| |
引用本文: | 张朝宗,肖永顺.对修订现行工业CT相关国家标准的建议[J].CT理论与应用研究,2015,24(4):629-635. |
| |
作者姓名: | 张朝宗 肖永顺 |
| |
作者单位: | 1. 清华大学核能与新能源技术研究院, 北京 100084; |
| |
摘 要: | 2012年以来我国发布了一系列关于工业CT的国家标准。虽然起到了相当的积极作用,但是仍然暴露出一些明显的不足之处。本文就名词术语标准的制定、空间分辨率的检测条件、对比度分辨率、低对比度空间分辨率、加强对于圆盘模体测定技术的研究和测定透射散射比等问题提出若干修订的建议。
|
关 键 词: | CT 空间分辨率 对比度分辨率 低对比度空间分辨率 圆盘模体 透射散射比 |
收稿时间: | 2015-05-20 |
Proposals of Revise on the Current National Standards of Industrial Computed Tomography |
| |
Institution: | 1. Institute of Nuclear and New Energy Technology, Tsinghua University, Beijing 100084, China;2. Department of Engineering Physics, Tsinghua University, Beijing 100084, China |
| |
Abstract: | Since 2012 a set of National Standards on Industrial Computed Tomography has been issued in China. Although they have taken some active effect, a lot of obvious deficiency is exposed. In this paper some proposals of revise about definition of technical terms, testing condition of spatial resolution, contrast resolution, low-contrast spatial resolution, study of testing using disk phantom, measure of transmission to scattering ratio etc. are presented. |
| |
Keywords: | computed tomography spatial resolution contrast resolution low-contrast spatial resolution disk phantom transmission to scattering ratio |
本文献已被 CNKI 万方数据 等数据库收录! |
| 点击此处可从《CT理论与应用研究》浏览原始摘要信息 |