首页 | 本学科首页   官方微博 | 高级检索  
     


Front‐ vs. back‐illuminated CCD cameras for photometric surveys: a noise budget analysis
Authors:N. Crouzet  T. Guillot  F. Fressin  A. Blazit  the A STEP team
Abstract:Exoplanetary transit and stellar oscillation surveys require a very high precision photometry. The instrumental noise has therefore to be minimized. First, we perform a semi‐analytical model of different noise sources. We show that the noise due the CCD electrodes can be overcome using a Gaussian PSF (Point Spread Function) of full width half maximum larger than 1.6 pixels. We also find that for a PSF size of a few pixels, the photometric aperture has to be at least 2.5 times larger than the PSF full width half maximum. Then, we compare a front‐ with a back‐illuminated CCD through a Monte‐Carlo simulation. Both cameras give the same results for a PSF full width half maximum larger than 1.5 pixels. All these simulations are applied to the A STEP (Antarctica Search for Transiting Extrasolar Planets) project. As a result, we choose a front‐illuminated camera for A STEP because of its better resolution and lower price, and we will use a PSF larger than 1.6 pixels. (© 2007 WILEY‐VCH Verlag GmbH & Co. KGaA, Weinheim)
Keywords:instrumentation: detectors  methods: numerica  technique: photometric
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号