首页 | 本学科首页   官方微博 | 高级检索  
     

改进消除温度对 CG-2型重力仪影响的装置/
引用本文:钱瘦石,辛程高. 改进消除温度对 CG-2型重力仪影响的装置/[J]. 地震学报, 1983, 5(3): 377-382
作者姓名:钱瘦石  辛程高
作者单位:1. 山东省地震局2. 山东潍坊农场重力站
摘    要:本文叙述采用改进后的双层恒温箱控制 CG-2型重力仪的工作环境的温度,对温度干扰问题做进一步的观测实验.在仪器室年温差仍为10℃左右变化的条件下,当恒温精度提高到25℃0.02℃时,消除了用恒温精度为24℃0.15℃的可控硅电热单层恒温所获得的资料所存在有一年为周期的,幅度可达500微伽的缓慢变化,并着重介绍了改进恒温装置的制作原理. 


IMPROVEMENT OF THE DEVICE FOR REDUCING TEMPERATURE EFFECT ON A CG-2 GRAVIMETER
Affiliation:mological Bureau of Shandong Province
Abstract:An improvement by using double thermostat for the device of reducing temperature effect on the CG-2 gravimeter has been described. Further tests were made to remove the slow yearly variation of amplitude 500 microgals of gravity by raising the precision of teme-prature control from 24℃ 0.15 in the case of single layer SCR thermostat to 25℃0.02, under the same condition of yearly room temperature change of 10℃. The paper gives also the constructional details of the device. 
Keywords:
本文献已被 CNKI 等数据库收录!
点击此处可从《地震学报》浏览原始摘要信息
点击此处可从《地震学报》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号