Comparison of Pb Purification by Anion-Exchange Resin Methods and Assessment of Long-Term Reproducibility of Th/U/Pb Ratio Measurements by Quadrupole ICP-MS |
| |
Authors: | Balz S Kamber Annette H Gladu |
| |
Institution: | Department of Earth Sciences, Laurentian University, 935 Ramsey Lake Road, Sudbury, ON P3E 2C6, Canada |
| |
Abstract: | A comparison between HBr-HCl and HBr-HNO3 based anion chemistry is presented to test the efficiency of Pb purification in the preparation of samples for isotope ratio measurement by ICP-MS. It was found that the small advantages in yield and blank offered by the HNO3-based method were more than compensated by the more effective matrix removal of the HCl-based method. Apart from very zinc rich matrices (e.g., sphalerite), a careful single pass purification using HBr and HCl removed more than 99.9% of the matrix. In preparation for the isotope ratio analysis, a small (2–5% m/v) liquid sample aliquot was analysed to determine U, Th and Pb concentrations by solution quadrupole ICP-MS. This allowed accurate prediction of the expected ion signal and permitted optimal spiking with Tl, if desired, for mass bias correction. Long-term results for international rock reference materials showed reproducibilities of better than 1% (Th/U) and 1.5% (U/Pb). For most geological applications, such analyses obviate the need for isotope dilution concentration measurements. |
| |
Keywords: | anion exchange Pb purification matrix removal |
|
|