Parallel calibration transfer and systematic effects in retrospective absorbed dose estimation using OSL
Institution:1. SUERC, University of Glasgow, Scotland, UK;2. LPSR, Instituto Superior Técnico, Universidade de Lisboa, Portugal;4. GeoBioTec, Universidade de Aveiro, Portugal;1. University of Tübingen, Department of Geosciences, Wilhelmstrasse 56, 72074, Tübingen, Germany;2. Technische Universität Bergakademie Freiberg, Geologie, Bernhard-von-Cotta-Strasse 2, 09599, Freiberg, Germany;3. Helmholtz Institute Freiberg for Resource Technologies, Division of Exploration, Halsbrücker Strasse 34, 09599, Freiberg, Germany;1. Centre for Earth Sciences, Indian Institute of Science, Bengaluru 560012, India;2. Centre for Ecological Sciences, Indian Institute of Science, Bengaluru 560012, India;3. Center for Chronological Research, Nagoya University, Furo-cho, Chikusa-ku, Nagoya 464-8602, Japan;4. Divecha Centre for Climate Change, Indian Institute of Science, Bengaluru 560012, India;1. Applied Nuclear Techniques in Geosciences Key Laboratory of Sichuan Province, Chengdu University of Technology, Chengdu, 610059, China;2. Sichuan Management & Monitoring Center Station of Radioactive Environment, Chengdu, 610031, China;1. Department of Psychiatry and Behavioral Sciences, Stanford University School of Medicine, Stanford, CA, USA;2. Sierra Pacific Mental Illness, Research, Education and Clinical Centers (MIRECC), Veterans Affairs Palo Alto Health Care System, Palo Alto, CA, USA;3. School of Psychology, University of Queensland, Brisbane, Australia;4. Geriatric Research, Education and Clinical Center (GRECC), Veterans Affairs Palo Alto Health Care System, Palo Alto, CA, USA;5. Department of Psychiatry, University of Pittsburgh, Pittsburgh, PA, USA
Abstract:Parallel multiple aliquot calibration transfer is combined with the use of standardized optically stimulated luminescence signals and evaluation of single aliquot regenerative dose response characteristics, to produce a robust and efficient transfer protocol for mineral samples used in dating and retrospective dosimetry. Transfer was made from an IST-LPSR 60Co primary air kerma standard to four Risø and Daybreak 90Sr/90Y irradiators, for quartz or polymineral grains of 90/100–160 μm or 160–250 μm diameter from seven samples. Grains were pretreated by activation, heating, or bleaching, and mounted on 0.5 mm thick aluminium or stainless steel disks, or 0.25 mm thick stainless steel cups. Multiple aliquot conversion coefficients based on parallel irradiations (mGy60Co per s of 90Sr/90Y exposure) were corrected for non linearity in dose response using the regenerative measurements, and compared with “single aliquot” coefficients obtained directly from the regenerative, i.e. retrospectively measured, dose responses. Ratios of measured/given β exposure time provide controls: these were close to unity for activated and heated material, which exhibited predose sensitization, and vice versa for optically bleached samples. Parallel multiple aliquot calibration transfer, using OSL integrated over the main signal decay, was found to offer better accuracy and precision than retrospective single aliquot measurements, and was robust for polyminerals as well as quartz. Differences in conversion coefficient between grain-sizes and supports on a given irradiator ranged up to 25% and were specific to the irradiator-support-grainsize permutation. Geometric effects are quantitatively explained by the solid angle subtended at the source by the sample, and the effect of support material (and thickness) by differences in electron backscatter.
Keywords:Calibration  OSL  Quartz/polymineral  SAR dose recovery  Precision  Accuracy
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号