Analysis of individual fluid inclusions using synchrotron X-ray fluorescence microprobe: progress toward calibration for trace elements
Authors:Bénédicte Ménez  Pascal PhilippotMichelle Bonnin-Mosbah  Alexandre SimionoviciFrançois Gibert
Institution:1 Laboratoire Pierre Süe, CEA-CNRS, Batiment 637, C. E. Saclay, 91191 Gif Sur Yvette cedex, France
2 Laboratoire de Géosciences Marines, CNRS-FRE 2316, Institut de Physique du Globe, case 89, T26-00 E3, 4 place Jussieu, 75252 Paris cedex 05, France
3 European Synchrotron Radiation Facility, μFID group, ID22, BP 220, 38043 Grenoble cedex, France
4 CNRS, UMR 6524, Université Blaise Pascal, 63038 Clermont Ferrand, France
Abstract:A critical problem for conducting quantitative analysis of individual fluid inclusions using Synchrotron X-Ray Fluorescence (SXRF) technique relates to the standardization and the calibration of the X-ray spectra. In this study, different approaches have been tested for calibration purposes: (a) the use of chlorine when Cl content can be estimated either from melting point depressions of undersaturated fluid inclusions or from saturation limits for halite-bearing fluid inclusions, (b) the use of calcium from synthetic fluid inclusions of known CaCl2 content as an external standard. SXRF analysis was performed on individual fluid inclusions from the Chivor and Guali emerald deposits, Columbia. These well-known samples contain a single fluid inclusion population for which detailed crush-leach analyses are available, thus providing a relevant compositional reference frame. Concentration estimates were also compared to Particle Induced X-ray Emission (PIXE) analysis carried out independently on the same fluid inclusions.Results of the calibration tests indicate that major (Cl, K, Ca, Fe, Mn) and trace element (Cu, Zn, As, Br, Rb, Sr, Ba, Pb) concentration estimates can be performed without precise knowledge of the analytical volume and the inclusion’s 3D geometry. Although the standard deviation of the SXRF results can be relatively high depending on the calibration mode used, mean concentration estimates for most elements are in good agreement with PIXE and crush-leach analysis. Elemental distributions within single fluid inclusions were also established. Associated correlation diagrams argue for the homogeneous distribution of most elements in the fluid inclusion. In contrast, Br shows a bimodal distribution interpreted to reflect a significant enrichment of the vapor portion of the inclusion fluid.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号