An In Situ Device to Measure Oxygen in the Vadose Zone and in Ground Water: Laboratory Testing and Field Evaluation |
| |
Abstract: | Oxygen probes developed to measure in situ oxygen concentrations in gaseous and aqueous environments were evaluated in laboratory tests and during long-term field evaluation trials at contaminated sites. The oxygen probes were shown to have a linear calibration and to be accurate compared to conventional dissolved oxygen electrodes and gas chromatography, both of which require labor-intensive sample collection and processing. The probes, once calibrated, required no maintenance or recalibration for up to a period of 7 years in low-oxygen environments, demonstrating long-term stability. Times to achieve 90% of the stabilized concentration ( t 90) after a step change in aqueous oxygen concentration were 100–120 min in laboratory experiments and up to 180 min in field experiments, which is adequate for monitoring subsurface changes. Field application data demonstrated that the oxygen probes could monitor oxygen concentrations in hydrocarbon-contaminated ground water to a depth of 20 m below the water table or in pyritic vadose zones over extended periods. During bioremediation field trials, oxygen monitoring enabled estimation of oxygen utilization rates by microorganisms and hydrocarbon biodegradation rates. Also, probes were able to monitor the development of ground water desaturation during air sparging trials, enabling the automated assessment of the distribution of injected air. |
| |
Keywords: | |
|
|