采用《测量不确定度评定与表示指南》,以等离子体发射光谱法测定土壤中的稀土元素铈为例,对测定结果进行不确定度评定。分析了不确定度的重要来源,包括溶液制备过程中引入的不确定度、样品称量引入的不确定度、标准物质外标法测量不确定度及仪器重复测定的不确定度。提供了引入不确定度各参数的采集和计算方法,对各不确定度分量进行分析计算,最后合成标准不确定度,通过乘以95%概率下的扩展因子2,获得测量结果的扩展不确定度。 相似文献
9.
10.
11.
电感耦合等离子体质谱法同时测定地下水中硼溴碘 总被引:2,自引:4,他引:2
建立了电感耦合等离子体质谱法同时测定地下水中B、Br、I的方法。选定φ=2%(体积分数)的稀NH3.H2O介质消除碘的记忆效应。采用干扰较少的10B和79Br同位素。B、Br、I在0~10 000 ng/mL呈良好的线性关系。方法的检出限为10B 0.176 ng/mL,79Br 0.876ng/mL,127I 0.132 ng/mL;精密度(RSD,n=12)为10B 2.86%,79Br 3.36%,127I 2.69%;10B的阶梯加标回收率为94.6%~101.5%,79Br为98.3%~104.9%,127I为96.5%~102.0%。 相似文献
12.
应用电感耦合等离子体质谱法(ICP-MS)测定地质样品中的痕量镉,存在多种质谱干扰,通常采用在线或离线方程进行校正,当样品中含锡较高时,采用传统固定系数校正方程,易导致测定结果有明显偏离,甚至结果出现负数。本文针对含高锡的地质样品,应用ICP-MS测定其中的镉,采用氢氟酸-高氯酸-硝酸敞开酸溶消解、硝酸浸提体系处理样品,通过测定~(111)Cd、~(113)Cd、~(114)Cd同位素,研究了干扰元素Sn、In、Zr、Mo对镉测定的影响。结果表明镉与干扰源浓度变化呈非简单的正相关性。①同质异位素Sn或In产生的干扰增值Δ(~(114)Cd/~(114)Sn或~(113)Cd/~(113)In)随干扰源浓度增大逐渐变大,~(114)Sn对~(114)Cd的干扰系数在0.0272~0.0222,~(113)In对~(113)Cd的干扰系数在0.0670~0.0412;②Zr和Mo在测定条件下形成氧化物和多原子复合离子物质对Cd均产生不同程度的质谱干扰。通过测定与样品中干扰源浓度相近的单一标准溶液产生的Cd干扰值,经在线修正干扰系数(γ),建立了精确的校正方程。该方法经标准物质验证,准确度高,相对标准偏差在6.57%~9.94%(n=7),方法检出限为0.03mg/kg,达到了地质分析检测要求,为高锡地质样品中的痕量镉分析提供了依据。 相似文献
13.
电感耦合等离子体质谱法测定地质样品中锗和镉的干扰及校正 总被引:5,自引:3,他引:2
采用HF-HClO4-HNO3分解样品,8 mol/L HNO3提取剂,外标法校准,电感耦合等离子体质谱法同时测定地球化学样品中锗和镉.选择103 Rh为内标,确定了最佳仪器参数,研究了Zn、Zr、Sn、Ce、Nd、Sm对Ge和Cd的干扰,试验选择质量数74 Ge和114 Cd 作为测定同位素,采用数学公式校正法校正了来自Nd和Sm的二次电离离子对 Ge 的干扰、Sn对Cd的同质异位素干扰.方法检出限(10s)分别为Ge 30 ng/L、Cd 15 ng/L,精密度(RSD,n=15)为Ge 1·35%、Cd1.47%.对多种国家一级标准物质进行分析验证,结果与标准值相符.方法适用于地质样品中微量锗和镉的测定. 相似文献
14.
15.
Method of Data Reduction and Uncertainty Estimation for Platinum-Group Element Data Using Inductively Coupled Plasma-Mass Spectrometry 总被引:1,自引:0,他引:1
A data reduction method is described for determining platinum-group element (PGE) abundances by inductively coupled plasma-mass spectrometry (ICP-MS) using external calibration or the method of standard addition. Gravimetric measurement of volumes, the analysis of reference materials and the use of procedural blanks were all used to minimise systematic errors. Internal standards were used to correct for instrument drift. A linear least squares regression model was used to calculate concentrations from drift-corrected counts per second (cps). Furthermore, mathematical manipulations also contribute to the uncertainty estimates of a procedure. Typical uncertainty estimate calculations for ICP-MS data manipulations involve: (1) Carrying standard deviations from the raw cps through the data reduction or (2) calculating a standard deviation from multiple final concentration calculations. It is demonstrated that method 2 may underestimate the uncertainty estimate of the calculated data. Methods 1 and 2 do not typically include an uncertainty estimate component from a regression model. As such models contribute to the uncertainty estimates affecting the calculated data, an uncertainty estimate component from the regression must be included in any final error calculations. Confidence intervals are used to account for uncertainty estimates from the regression model. These confidence intervals are simpler to calculate than uncertainty estimates from method 1, for example. The data reduction and uncertainty estimation method described here addresses problems of reporting PGE data from an article in the literature and addresses both precision and accuracy. The method can be applied to any analytical technique where drift corrections or regression models are used. 相似文献
16.
17.
硅钼黄分光光度法测定地下水中偏硅酸的不确定度评定 总被引:2,自引:1,他引:2
采用不确定度连续传递模型,对硅钼黄分光光度法测定地下水中偏硅酸(DZ/T0064.62—93)的不确定度进行评定。测量结果的不确定度主要来源于标准溶液引入的不确定度、曲线拟合产生的不确定度和测量过程引入的不确定度三部分,而二氧化硅和偏硅酸的摩尔质量不确定度较小,可以忽略不计。采用双误差回归的方式对标准曲线进行拟合,在对各个不确定度分量进行量化的基础上,通过合成得到测量结果的标准不确定度,再乘以95%置信概率下的扩展因子2,得到测量结果的扩展不确定度。 相似文献
18.
敞口酸溶-电感耦合等离子体质谱法同时测定地质样品中45个元素 总被引:5,自引:12,他引:5
改进了三酸(HCl-HNO3-HF)或四酸(HCl-HNO3-HF-HClO4)溶矿对难溶元素分解不彻底及对易挥发元素测定结果不稳定的现象,建立了地质样品中15个稀土元素及其常规微量元素一次溶矿、一次测定的分析方法,提高了分析质量。采用HCl-HNO3-HF-HClO4-H2SO4混合酸一次分解样品,用50%的逆王水复溶提取后定容,以Rh为内标元素,三通在线进样方式,电感耦合等离子体质谱法同时测定地质样品中45个元素的含量。讨论了分解方法对易挥发元素、难溶元素的影响,对岩石标准物质(GBW07106)、水系沉积物标准物质(GBW07303)和土壤标准物质(GBW07429)的测定值与标准值相一致,方法精密度为0.03%~4.97%。方法准确度高,简单、快捷、实用。 相似文献