高频燃烧-红外吸收法用于分析矿石中低含量硫的测定结果较为准确,但对于高含量的硫,分析结果的准确度不高。本文采用高频燃烧-红外碳硫分析仪测定钼矿石和镍矿石中的高含量硫,选择纯铁屑和钨粒作助熔剂,高温燃烧分解样品,通过实验优化了样品称样量、助熔剂用量、仪器分析时间等测定条件。用国家标准物质进行验证,方法精密度(RSD,n=9)小于1%,加标回收率为96.0%~101.9%;与传统的硫酸钡重量法进行比对试验,测定值的相对误差小于2%。针对不同的矿石样品,研究了实际样品与标准物质的基体匹配问题,消除了基体效应的影响,对于钼矿石和镍矿石样品中含量在1%~35%范围内的硫,均能够准确测定,解决了钼矿石和镍矿石中高含量硫的快速、准确测定问题。  相似文献   

2.
应用高频燃烧-红外碳硫吸收法测定含铜烧结物中的高含量硫。对样品称样量、助熔剂的种类、加入顺序、用量等测定条件进行研究。由于样品含硫量的不同,根据矿石标准样品中含硫量的比例关系确定具体的称样量。以纯铁屑和钨粒作助熔剂,高温燃烧分解试样,红外检测可定量测定含铜烧结物样品中含量为1%~23%的硫。方法加标回收率为94%~113%,相对标准偏差(RSD,n=6)小于2%。与经典的硫酸钡重量法对比,相对误差小于4%。建立的方法解决了大宗含铜烧结物中硫快速、准确测定的问题,已应用于实际的日常检验工作。  相似文献   

3.
高频燃烧—红外碳硫仪测定地质样品中的碳和硫   总被引:13,自引:6,他引:7  
应用HIR-944B型高频-红外碳硫分析仪,对不同地质样品中碳、硫的测定进行了研究,称样30~60mg,加入0.4g纯铁屑及1.7g钨粒助熔剂,高温燃烧分解试样,红外检测,可定量地质样品中质量为0~O.9mg的硫及质量为0~15mg的碳.用该仪器测定地质标样中碳、硫的结果与标准值符合,碳和硫11次测定的RSD分别是<2.6%和<3.0%.  相似文献   

4.
红外碳硫仪测定矿石中高含量硫的探讨   总被引:4,自引:3,他引:1  
探讨了用红外碳硫测定仪测定矿石中高含量硫需注意的一些影响因素,包括标样的选择、减少天平误差、称样量、样品的充分燃烧以及仪器的稳定性,说明了用红外碳硫仪测定矿石中高含量硫的可行性。  相似文献   

5.
应用高频燃烧-红外碳硫分析仪测定农用地土壤质量调查样品中碳、硫的含量,解决了土壤样品中碳、硫测定过程中存在的基体影响和低电磁感应等问题。对样品称样量、助熔剂的加入量、氧气流量等测定条件进行优化试验,用国家一级标准物质验证了方法准确度和精密度,相对误差(RE)小于10%,相对标准偏差(RSD,n=8)总碳小于2%,硫小于9%。方法能够满足农用地土壤质量调查样品的分析质量要求。  相似文献   

6.
高频红外碳硫仪测定重晶石和黄铁矿中的硫   总被引:2,自引:2,他引:0  
硫酸盐矿石和硫化物矿石大部分是低电磁性的物质,利用高频红外碳硫仪测定这类矿石中的硫时,在燃烧过程中难以产生较大的电磁感应涡流,导致矿石中的硫释放不完全,造成硫的测定结果偏低。本文采用二氧化硅将重晶石精矿和黄铁矿精矿稀释成不同硫含量的重晶石和黄铁矿样品,通过优化称样量及助熔剂等测试条件,建立了使用高频红外碳硫仪测定重晶石和黄铁矿中硫含量的分析方法。结果表明:当样品中的硫含量高于2%时确定称样量为0.07 g,当硫含量低于2%时确定称样量为0.1 g,加入助熔剂0.4 g锡粒+0.4 g铁粒+1.5 g钨粒,可使重晶石和黄铁矿中的硫完全释放进入仪器红外吸收区域,硫的回收率提高至95.8%~104.2%(重晶石)和95.3%~105.1%(黄铁矿),分别高于常规红外碳硫仪的回收率(83.39%~91.1%和91.5~97.5%)。本方法精密度高(RSD5%),实现了硫含量的准确测定。  相似文献   

7.
测试地质样品中的硫含量,以电感耦合等离子体发射光谱法(ICP-OES)和燃烧-红外吸收光谱法应用最为广泛。ICP-OES法灵敏度高、稳定性好,但受样品预处理和基体干扰的影响较大;燃烧-红外吸收光谱法便捷高效,但受结晶水红外吸收干扰,分析硫含量低的样品稳定性较差。本文采用5种酸溶方式处理样品ICP-OES测定硫含量,同时采用燃烧-红外吸收光谱法测定低中高含量的硫,综合比较了两类方法的检出限、检测范围、精密度和准确度、分析效率等,明确了各方法的适用范围。实验中确定了燃烧-红外吸收光谱法最佳测试条件为:称样量0.0500g,燃烧时间25s,分析时间40s,氧气流量4.0L/min;通过标准物质验证,该方法检出限为10×10-6,检测范围为10×10-6~470000×10-6,相对标准偏差(RSD) < 6%(n=12),相对误差绝对值小于8%。实验结果表明,ICP-OES分析效率高,但是样品处理时间长,检测范围不如燃烧-红外吸收光谱法宽;燃烧-红外吸收光谱法采用固体直接进样,成本低,用高氯酸镁作为干燥剂可解决结晶水红外吸收干扰问题。总体上,ICP-OES法适用于分析硫含量低的样品或作为测试结果佐证的手段,可实现多元素联测;批量样品或基体类型复杂的样品宜采用燃烧-红外吸收光谱法测试,更加便捷。  相似文献   

8.
为提高土壤及水系沉积物中全硫含量测定结果准确度,提出了库仑滴定法进行全硫量测定的方法优化,探究了称样量、炉温、载气流量和催化剂种类等不同因素影响对土壤及用水系沉积物中硫含量的测定准确度的影响。测试结果表明:库仑滴定法测定土壤及水系沉积物中硫含量的最佳工作条件为炉温1 200℃,称样量0.100 0g,载气流量1.0 L/min;纯铁催化剂对于极高或极低硫值样品的测定结果稳定度更高,WO3催化剂具有更高的性价比;库仑滴定法相较于燃烧-碘量法和高频红外法对低硫含量样品测定更准确。相较于燃烧-碘量法和高频红外法,库仑滴定法对低硫含量样品测定更准确。  相似文献   

9.
耶曼  李婧  马怡飞  柯艳  李小桂 《岩矿测试》2022,41(4):680-687
高频红外碳硫仪测定不同矿石种类中的硫含量,测定速度快,稳定性好,但当测定范围宽、样品种类多时,受助熔剂、氧化温度和氧化时间影响较大。本文应用高频红外碳硫分析仪,研究了实验条件对分析结果的影响,通过优化样品称样量、助熔剂添加量和分析时间,建立了矿石样品中质量分数为0.74%~32.0%的硫含量检测方法,分析条件为:分析氧气流速2.8L/min,样品称样量0.0400g,纯铁助熔剂0.50g,纯钨助熔剂2.0g,分析时间45s。通过国家标准物质验证该方法的检出限为0.185%,定量限为0.739%,标准曲线线性相关系数大于0.9995,测定结果的相对标准偏差小于3%(n=11),与标准值的相对误差小于2%,且均小于DZ/T 0130—2006中对矿石样品分析要求的相对误差允许限。采用本方法与传统燃烧碘量法对实际样品进行测定,两种方法测定值的绝对误差小于0.5%,测定结果之间呈极显著线性关系(R2=0.9995),表明两种方法具有良好的一致性。  相似文献   

10.
红外碳硫分析仪测定土壤样品中的碳和硫   总被引:1,自引:0,他引:1  
刘志  段英楠 《吉林地质》2009,28(3):103-105
应用高频一红外碳硫分析仪,进行了土壤样品中碳、硫的测定,用该仪器测定土壤标准物质中碳、硫的结果与标准值符合。碳和硫11次测定的RSD分别是〈3%和〈10%。  相似文献   

11.
A study has been undertaken to determine sulfur in geological samples by coupled analytical techniques. Two measurement methods have been developed: one using an electric furnace coupled to an ion chromatograph (electric furnace-IC) and another using infrared (IR) and quadrupole mass spectrometry (QMS) for evolved gas analysis (EGA) coupled with a thermogravimetric analyser (TGA). In the electric furnace-IC method, measurement was performed without any sample pre-treatment. The measurement conditions were optimised by varying sample quantity, type of catalyst (WO3, Cu, W and V2O5) and sample/catalyst ratio, and the detection limit was 10 μg g−1. Sulfur ores decompose at different temperatures. However, TGA-EGA allowed identification of the different forms of sulfur in the sample, even when they were found in very low concentrations, because the sulfur was continuously analysed. The developed chromatographic method allowed simultaneous analysis of several sample components, such as S, Cl and F, with a low detection limit. The method was much faster and more specific than the methods described in the literature. The results of the sulfur determination had low scatter, possibly because the samples underwent little handling during analysis: the operator only weighed and placed the sample in the furnace, the rest of the measurement process was fully automated. The results obtained by both the developed methods have been validated by using reference materials and comparison with combustion-IR spectroscopy, a standard method for determining total sulfur in a sample.  相似文献   

12.
Total carbon and sulfur were determined on a variety of geochemical reference samples such as rocks, sediments, minerals and ores by a method of infrared absorption following combustion in a high-frequency induction furnace. The sample was burned with an accelerator of iron powder and tungsten chips, and the carbon and sulfur dioxide in the evolved gases have been determined directly by a simultaneous infrared analyzer. The agreement between present results and other data is generally good.  相似文献   

13.
高炉喷吹用煤可减少焦炭的使用量,是现代高炉降低生产成本提高经济效益的重要技术措施。从高炉喷吹用煤的煤质影响因素入手,在稀缺煤和优质煤煤质评价的基础上探讨稀缺高炉喷吹用煤的地质评价指标及分级参数。通过分析,将高炉喷吹原料用煤的煤质评价指标划分为工艺指标和经济指标两大类,工艺指标是决定煤是否能用于高炉喷吹工艺的关键指标,经济指标主要影响高炉喷吹工艺的成本。煤类差异是高炉喷吹用煤选择的首要因素,根据我国煤炭资源禀赋特征,提出无烟煤、贫煤、贫瘦煤是高炉喷吹用煤中的稀缺煤类。结合煤炭地质勘查中对\  相似文献   

14.
不同景观条件下选取合适的采样粒级是准确地获取地球化学信息,提升地质找矿效果的关键。为研究中国南部湿润半湿润中低山丘陵景观区水系沉积物测量最佳采样粒级,本文在安徽省南部胡乐司—宁国墩地区开展了区域化探采样粒度方法技术试验,采集-60目和-10目~+80目两种粒级样品,采用波长色散X射线荧光光谱法和电感耦合等离子体质谱法为主体的配套分析方案和测试技术获取了全国区域化探扫面规定的40种元素高精度数据。研究表明:两种粒级水系沉积物中造岩元素背景含量与区域岩石背景值基本接近,其他微量元素多呈富集状态,以-10~+80目粒级中As、Au、Hg、Mo、Pb、Sb等成矿元素富集程度最强;-10~+80目粒级受后期表生改造作用影响较弱,最大限度保留了原生地球化学分布特征,并在准确圈定与矿化有关的异常和清晰反映矿致异常特征等方面明显优于-60目粒级。因此,建议在安徽省南部中低山丘陵景观区,选择-10~+80目粒级为水系沉积物测量的最佳采样粒级,研究结果可以为中国相似景观区地球化学勘查找矿水系沉积物测量采样粒级的选择提供参考。  相似文献   

15.
微量陨石激光熔样稀有气体测定方法是一种可以在微米尺度上对几毫克陨石样品进行准确稀有气体同位素分析的方法,克服了传统全岩熔融法在测量时存在样品用量大、前处理过程复杂和样品稀有气体分布不均导致不同组分的宇宙射线暴露历史无法进一步区分等问题。但是由于该方法所用样品体积小和样品用量低,要求实验室具有超低本底的稀有气体提取系统,目前国内在微量陨石稀有气体分析技术方面尚处于起步阶段。本文采用金刚石激光样品窗成功研制了超低本底的气体提取系统,通过系统体积标定和天平称量误差、热本底、干扰元素、质量歧视及质谱灵敏度等参数的校正,在中国科学院地质与地球物理研究所建立了微量陨石激光熔样稀有气体测定方法,并对毫克级微量钙长辉长无球粒陨石Millbillillie粉末标样进行了稀有气体同位素含量和比值测定,计算获得准确一致的宇宙暴露年龄。该方法的建立,将为我国迅速发展的比较行星学和深空探测提供重要技术支撑。  相似文献   

16.
探针原子化石墨炉原子吸收法测定高纯金属镁中痕量铅   总被引:2,自引:2,他引:0  
使用自制的全自动探针原子化装置,将探针原子化技术应用于测定高纯金属镁中痕量铅,利用样品中的基体镁转化成的硝酸镁作为基体改进剂,进一步改善了铅的分析性能。通过实验,确定了最佳的仪器条件和探针原子化分析条件,并与管壁原子化法进行了比较。方法检出限为4.3×10-12gPb,相对标准偏差(RSD,n=6)为8.2%,高纯金属镁中加标回收率在93%~104%。  相似文献   

17.
采用大试管和红外加热的多孔控温消解炉的简单装置消解植物样品,电感耦合等离子体发射光谱/质谱法测定其中的主量和痕量元素,对消解温度、消解时间、消解溶剂、样品取样量等条件进行考察。用植物国家一级标准物质进行前处理及测定,验证方法的可行性,测定值与标准值基本吻合。方法不仅能同时测定B、Ba、Ca、Cd、Cu、K、Mg、Mn、Na、Ni、P、S、Zn、CoSr、Pb、Cr等元素,还适用于Hg、As挥发性元素的消解测定。方法的精密度(RSD,n=10)≤10%。  相似文献   