Comparison of Image Quality between Virtual Monochromatic Images and Conventional CT Images
-
摘要: 目的:对比双源双能CT的虚拟单能图像与常规CT图像的图像质量。方法:采用西门子SOMATOM Force双源双能CT扫描图像质量控制体模Catphan 500的高对比度分辨率CTP 528组件和低对比度分辨率CTP 515组件,分别采用双球管扫描方案重建低、中、高3档虚拟单能能谱图像(A组:40 keV,B组:70 keV,C组:100 keV)和一种单球管扫描方案图像(D组:120 kVp)。两名放射科医师分别测量CTP 528的高对比分辨率(最佳线对数)和CTP 515的低对比度分辨率(1% 对比度浓度下的最小孔径);随后测量各组图像的平均噪声、信噪比(SNR)及对比噪声比(CNR)。结果:两名医师的图像一致性评价良好(高/低对比度分辨率:Kappa系数为0.667;噪声参数:Spearman相关系数r=0.920。各组图像中,D组高对比度分辨率最佳(线对6 lp/cm);A、B组图像低对比度分辨率最佳(孔径均为4 mm)。C组图像平均噪声最低(2.333±0.289),SNR表现最佳(37.944±3.949);噪声和SNR在各组图像间差异有统计学意义。B组图像CNR最高(3.361±0.875),但各组图像间CNR差异不具有统计学意义。结论:双能CT虚拟单能图像相比常规CT图像可以在保持低对比度分辨率的基础上,改善图像噪声;随着虚拟单能能级的增加,图像噪声减小,信噪比增加。Abstract: Objective: To compare the image quality between virtual monochromatic images (VMI) of dual-source dual-energy CT (DECT) and conventional CT images. Siemens SOMATOM Force dual source dual energy CT was adopted in the scanning of the high contrast resolution CTP 528 component and low contrast resolution CTP 515 component of image quality control phantom catphan 500. Dual-tube dual-energy scanning scheme was used to reconstruct low, medium and high-level VMI (group A: 40 keV, group B: 70 keV, group C: 100 keV) and a single tube scanning scheme was used to reconstruct conventional images (Group D: 120 kVp). Two radiologiss carried out the measurement of high resolution by the smallest line pair one could distinguish and low-contrast sensitivity was measured by the smallest diameter under 1% contrast. Mean noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were also evaluated for each group. Results: The image consistency of the two radiologists turned out well (high reolution and low-contrast sensitivity: Kappa coefficient=0.667; noise parameter: Spearman r = 0.920). Among all images, D group performed the best in high resolution (6 lp/cm) and A and B groups showed the best low-contrast sensitivity (diameter of 4 mm). C group had the lowest noise (2.333±0.289) and highest SNR (37.944±3.949). Mean noise and SNR were found significantly difference among the four groups. B group showed the highest CNR (3.361±0.875), but no statistical significance was found among the groups. Conclusion: Compared with conventional CT, VMI images from DECT could improve image quality while retain low-contrast sensitivity; image noise decreases while SNR increases as VMI elevates.
-
Keywords:
- dual-energy CT /
- virtual monochromatic image /
- phantom
-
常规CT图像实际上是在混合能量X线束的基础上成像,低能X线束易被人体组织吸收,而透过的高能X线束容易产生射线束硬化伪影,现在通常的做法是采用特定的算法对射线束硬化伪影进行校正。但是由于常规CT本身的缺陷,这种校正常常不能充分对其进行抵消,随之产生的非线性伪影会造成疾病的误诊和漏诊[1-2]。同时,由于光电效应和康普顿散射的存在,拥有不同原子序数的物质会呈现出相似的表现,使得CT的密度分辨率不佳[3]。此外,如果采用低电压扫描图像噪声会增加,但图像的对比度分辨率更好。而采用高电压方案则图像的噪声得到抑制,但对比度分辨率进一步下降,且同等情况下的辐射剂量增加。
双能CT利用高低两种能量进行成像,在此基础上通过物质分离算法,重建的虚拟单能图像可以进一步模拟各级能量下不同物质的衰减[4-5]。既往研究表明,双能CT虚拟单能成像相比传统CT可以降低图像噪声,提升图像质量[6-10]。但是,相关研究大多都是临床研究,受限于辐射剂量,不能做到同一个研究对象的自身对照,患者差异会是影响研究结果的潜在重要影响因素。
因此,本研究旨在运用体模,对比双源双能CT低、中、高3档虚拟单能能级图像和两种常规CT图像的图像质量,探讨虚拟单能图像的临床可行性。
1. 资料与方法
1.1 体模
采用CT图像质量控制体模Catphan 500中的高对比度分辨率CTP 528和低对比度分辨率CTP 515两个组间。CTP 528(图1(a))由21组高密度线对(呈放射状分布)的结构,分辨率为1~21 lp/cm。CTP 515(图1(b))由内外两层低密度孔阵(呈放射状分布)的结构构成,内层孔阵的对比度分别为0.3%、0.5%和1.0%,直径分别为3、5、7和9 mm;外层孔阵的对比度分别为0.3%、0.5%和1.0%,直径分别为2、3、4、5、6、7、8、9和15 mm。
1.2 仪器与方法
通过西门子SOMATOM Force双源双能CT分别扫描Catphan 500的CTP 528和CTP 515组件(图1),分别采用双球管双能模式扫描与单球管单能模式扫描。所有扫描方案固定剂量20 mGy,螺距1.0,机架旋转速度为1 s/r。其中双能模式扫描参数为双球管电压对100/150 Sn kVp,电流294 mAs/147 mAs;单能扫描方案采用120 kVp,电流分别为299 mAs和205 mAs。
所有图像重建层厚5 mm,标准滤波函数Qr40(双能模式)或Br40(单能模式)和中等迭代重建算法ADMIRE 2。双能扫描重建3档虚拟单能图像:A组40 keV,B组70 keV,C组100 keV;单能混合能量扫描图像为:D组120 kVp。
1.3 图像质量评估
评估CT图像质量的几个主要参数分别为:高对比度分辨率、低对比度分辨率、图像噪声、信噪比(signal-to-noise ratio,SNR)和对比噪声比(contrast-to-noise ratio,CNR)。
所有图像由两名具有5年以上阅片经验的放射科医生采用独立盲法测量并分别进行主观与客观评价。主观评价记录包含高对比度分辨率组件CTP 528的最佳线对(1~2 lp/cm)以及低对比度分辨率组件CTP 515所能分辨的外层1% 浓度下的最小孔径(直径2~15 mm,要求分辨80% 以上有效面积),比较两名医生的主观评价结果,如有分歧则由另一名高年资医生做最终评判。
客观测量选取两个面积均为100 mm2的圆形感兴趣区域(region of interest,ROI)分别测量CTP 515组件中1% 浓度中直径最大(15 mm)孔径和邻近固态水背景的CT值(CTm和CTw,单位HU),将15 mm孔径ROI标准差(SD)定义为噪声(单位HU),随后计算各组图像SNR=CTm/SD和CNR=(CTm- CTw)/SD。所有主观及客观评价指标均在各组件中心的连续3个层面上进行测量;主观指标取3个层面的最佳值。
1.4 统计学分析
运用SPSS 23.0统计分析软件进行分析。连续性变量数据以
$\bar{x}\pm s $ 表示。两名医生的主观一致性评价采用统计分析Cohen's Kappa检验,客观一致性评价采用Spearman相关性分析。5组图像的高对比度分辨率和低对比度分辨率采用秩和检验;5组图像的SD、SNR及CNR的比较采用单因素方差分析。双能模式虚拟能级图像的组内比较采用Bonferroni矫正,以P<0.05认为差异具有统计学意义。
2. 结果
2.1 一致性评价
两名医生对各组图像的高对比度分辨率和低对比度分辨率的主观及客观一致性评价良好,Kappa系数为0.667,Spearman相关系数r=0.920,差异均具有统计学意义。
2.2 四组图像对比度分辨率比较
D组图像具有最佳的高对比度分辨率(线对6 lp/cm),A组~C组图像的最佳线对数均为5 lp/cm。对于低对比度分辨率,A组和B组图像表现最佳(均为4 mm孔径),其余两组图像的最小孔径为5 mm(表1)。3档虚拟单能图像对比见图2。
表 1 四组图像高对比度分辨率和低对比度分辨率比较Table 1. Comparison of high-contrast resolution and low-contrast resolution of four groups of images对比度分辨率 A组
(40 keV)B组
(70 keV)C组
(100 keV)D组
(120 kVp)高对比对分辨率(线对/(lp/cm)) 5 5 5 6 低对比度分辨率(孔径直径/mm) 4 4 5 5 2.3 四组图像噪声、SNR和CNR比较
各组图像间噪声差异有统计学意义,其中A组图像噪声最大(4.167±0.289),C组图像噪声最小(2.333±0.289);噪声在虚拟单能组内(A组~C组)差异有统计学意义。各组图像间SNR差异有统计学意义,C组图像SNR最高(37.944±3.949);SNR随着虚拟单能能级的升高而增加,SNR在虚拟单能组内(A组~C组)间差异有统计学意义。B组图像的CNR最高(3.361±0.875),但各组图像CNR差异无统计学意义;组内比较方面,虚拟单能组(A组~C组)的CNR差异也无统计学意义(表2)。
表 2 四组图像图噪声参数评价结果比较Table 2. Comparison of the evaluation results of the noise parameters of the five groups of images组别 SD/HU SNR CNR A组(40 keV) 4.167±0.289 10.942±0.465 2.383±0.336 B组(70 keV) 2.500±0.500 24.028±5.000 3.361±0.875 C组(100 keV) 2.333±0.289 37.944±3.949 3.139±0.337 D组(120 kVp) 2.500±0.000 30.806±1.398 2.778±0.241 $P $ <0.001 <0.001 0.179 3. 讨论
CT在临床诊疗中扮演着重要的作用,但常规CT在选择最优管电压方面具有一定的妥协性。选择低电压代表更低的辐射剂量,虽然其具有更高的图像对比度,在常规低剂量肺筛的作用显著,但是也在一定程度上牺牲了图像质量,增加了图像噪声,因此在其他领域的作用受限;而选择高电压方案则与此相反,虽能获得更佳的图像质量,但其图像对比度下降且会增加病人的受照辐射剂量。
由于常规CT扫描采用的是混合能量X射线谱,以120 kVp为例,低能的X射线被组织吸收,导致其平均能量大约在90 keV左右,在这种能量下,钙与碘具有相似的衰减系数,因此两种物质在常规CT上具有相似的高密度,这不利于钙化斑块的检出[11]。双能CT在不增加患者辐射剂量的基础上,可以重建不同虚拟单能能级的图像,利用不同物质/组织在不同X线束下的衰减系数不同,以此获得不同对比度的CT图像,并且可以通过“动态”观察不同物质随着能级变化的特点,实现常规CT上不能实现的微小病灶的检出和病灶成分的鉴别。目前有研究证明其在心血管疾病、肿瘤检出、颅内出血、炎症性肠病等领域的效果显著[12-16]。
本研究分别研究低、中、高3档虚拟单能能级图像的图像质量,3档能级图像具有相似的高/低对比度分辨率,100 keV低对比度分辨率稍有下降。既往研究表明,低keV下图像的低对比度分辨率(即密度分辨率)较高,图像对不同软组织的分辨较好,但是同时伴随着一定的伪影,而高keV下图像的伪影得到抑制,但图像的密度分辨率下降。因此,低keV虚拟单能图像有助于CT静脉造影及软组织肿块等的检出,而高keV则有助于抑制如心脏支架、人工髋关节伪影等对图像质量的影响[4,11]。
图像噪声方面,随着能级的增加,图像平均噪声降低,而70 keV的CNR更高,100 keV的SNR最高。总体来看,本体模研究表明,70 keV是比较均衡的虚拟单能能级;不同研究也表明,70 keV有助于冠状动脉、肺血管、胰腺癌、小肠梗阻等的显示或检出[17-20]。目前最常用的3档双能成像模式(双源双能量、快速kVp切换以及双层探测器模式),与常规120 kVp图像等效的虚拟单能图像的能级均在70 keV左右。同等条件下,70 keV图像往往具有更低的辐射剂量,更佳的图像质量。本研究中,70 keV图像的平均噪声比常规120 kVp图像更低,而CNR更高,两者的高/低对比度分辨率相近,提示70 keV图像具有更高的图像质量,这与前人的研究结果相一致[21-22]。
本研究也存在一定的局限性。①本研究只对比了低、中、高3档虚拟单能能级,在全面对比虚拟单能图像之间的差异方面还显不足;②本研究用的体模为研究对象,且只选取了其中两个模块进行评价,对图像质量的全面评价略显不足,例如对比虚拟单能图像和常规CT图像的CT值准确性等方面还有待进一步研究;③本模型研究在临床方面的指导作用还有待进一步阐明。
总之,能谱CT虚拟单能图像相比传统混合能量CT图像可以降低图像噪声并提升图像质量。
-
表 1 四组图像高对比度分辨率和低对比度分辨率比较
Table 1 Comparison of high-contrast resolution and low-contrast resolution of four groups of images
对比度分辨率 A组
(40 keV)B组
(70 keV)C组
(100 keV)D组
(120 kVp)高对比对分辨率(线对/(lp/cm)) 5 5 5 6 低对比度分辨率(孔径直径/mm) 4 4 5 5 表 2 四组图像图噪声参数评价结果比较
Table 2 Comparison of the evaluation results of the noise parameters of the five groups of images
组别 SD/HU SNR CNR A组(40 keV) 4.167±0.289 10.942±0.465 2.383±0.336 B组(70 keV) 2.500±0.500 24.028±5.000 3.361±0.875 C组(100 keV) 2.333±0.289 37.944±3.949 3.139±0.337 D组(120 kVp) 2.500±0.000 30.806±1.398 2.778±0.241 $P $ <0.001 <0.001 0.179 -
[1] SUN H, QIU S, LOU S, et al. A correction method for nonlinear artifacts in CT imaging[C]//IEEE Engineering in Medicine and Biology Society. Conference Proceedings 2004, 2004: 1290-1293.
[2] RODRÍGUEZ-GRANILLO G A, ROSALES M A, DEGROSSI E, et al. Signal density of left ventricular myocardial segments and impact of beam hardening artifact: Implications for myocardial perfusion assessment by multidetector CT coronary angiography[J]. International Journal of Cardiovascular Imaging, 2010, 26(3): 345−354. doi: 10.1007/s10554-009-9531-5
[3] 任国庆, 滑炎卿, 李剑颖. CT能谱成像的基本原理及临床应用[J]. 国际医学放射学杂志, 2011,34(6): 559−563. doi: 10.3784/j.issn.1674-1897.2011.06.Z0612 REN G Q, HUA Y Q, LI J Y. The basic principle and clinical application of CT spectrum. imaging[J]. International Journal of Medical Radiology, 2011, 34(6): 559−563. (in Chinese). doi: 10.3784/j.issn.1674-1897.2011.06.Z0612
[4] Mc COLLOUGH C H, LENG S, YU L, et al. Dual- and multi-energy CT: Principles, technical approaches, and clinical applications[J]. Radiology, 2015, 276(3): 637−653. doi: 10.1148/radiol.2015142631
[5] 王文杰, 陈平, 潘晋孝, 等. 基于参考组分的双能CT成像方法[J]. CT理论与应用研究, 2021,30(1): 61−69. DOI: 10.15953/j.1004-4140.2021.30.01.06. WANG W J, CHEN P, PAN J X, et al. Dual-energy CT imaging method based on reference components[J]. CT Theory and Applications, 2021, 30(1): 61−69. DOI: 10.15953/j.1004-4140.2021.30.01.06. (in Chinese).
[6] 许伟雄, 黄旭璇, 林顺发, 等. 腹部动脉能谱CT混合能量与单能量图像的比较研究[J]. 中国CT和MRI杂志, 2014,12(8): 91−94. doi: 10.3969/j.issn.1672-5131.2014.08.30 XU W Q, HUANG X X, LIN S F, et al. Comparative study of abdominal artery energy spectrum CT mixed energy and single energy image[J]. Chinese Journal of CT and MRI, 2014, 12(8): 91−94. (in Chinese). doi: 10.3969/j.issn.1672-5131.2014.08.30
[7] 朱晓红, 刘斌, 周勇, 等. 冠状动脉能谱CT单能量成像与混合能量成像质量的比较−静态模型的实验研究[J]. 临床放射学杂志, 2012,31(5): 738−742. ZHU X H, LIU B, ZHOU Y, et al. Comparison of the quality of coronary artery energy spectrum CT single-energy imaging and mixed-energy imaging−An experimental study of a static model[J]. Journal of Clinical Radiology, 2012, 31(5): 738−742. (in Chinese).
[8] 韦炜, 邓克学, 赵英明, 等. 腹部双能量能谱CT成像中混合能量模式与单能量模式重建图像的比较研究[J]. 安徽医科大学学报, 2016,51(11): 1650−1653. WEI W, DENG K X, ZHAO Y M, et al. A comparative study of mixed energy mode and single energy mode in dual-energy spectrum CT imaging of the abdomen[J]. Journal of Anhui Medical University, 2016, 51(11): 1650−1653. (in Chinese).
[9] 李林坤, 王秋燕, 丛湘华, 等. 双下肢单能量60 keV与混合能量120 kVp CTA应用价值比较[J]. 临床放射学杂志, 2019,38(11): 2180−2184. LI L K, WANG Q Y, CONG X H, et al. Comparison of the application value of single energy 60 keV and mixed energy 120 kVp CTA for both lower limbs[J]. Journal of Clinical Radiology, 2019, 38(11): 2180−2184. (in Chinese).
[10] 王伟新, 张秋奂, 郭鹏德, 等. 能谱CT单能量结合ASiR技术对腹部静脉成像质量的优化研究[J]. CT理论与应用研究, 2019,28(1): 61−72. DOI: 10.15953/j.1004-4140.2019.28.01.07. WANG W X, ZHANG Q H, GUO P D, et al. Study on the optimization of energy spectrum CT single energy combined with ASiR technology on the quality of abdominal vein imaging[J]. CT Theory and Applications, 2019, 28(1): 61−72. DOI: 10.15953/j.1004-4140.2019.28.01.07. (in Chinese).
[11] ALBRECHT M H, VOGL T J, MARTIN S S, et al. Review of clinical applications for virtual monoenergetic dual-energy CT[J]. Radiology, 2019, 293(2): 260−271. doi: 10.1148/radiol.2019182297
[12] LEITHNER D, MAHMOUDI S, WICHMANN J L, et al. Evaluation of virtual monoenergetic imaging algorithms for dual-energy carotid and intracerebral CT angiography: Effects on image quality, artefacts and diagnostic performance for the detection of stenosis[J]. European Journal Radiology, 2018, 99: 111−117. doi: 10.1016/j.ejrad.2017.12.024
[13] KAUP M, SCHOLTZ J E, ENGLER A, et al. Dual-energy computed tomography virtual monoenergetic imaging of lung cancer: Assessment of optimal energy levels[J]. Journal of Computer Assisted Tomography, 2016, 40(1): 80−85. doi: 10.1097/RCT.0000000000000319
[14] MARTIN S S, WICHMANN J L, PFEIFER S, et al. Impact of noise-optimized virtual monoenergetic dual-energy computed tomography on image quality in patients with renal cell carcinoma[J]. European Journal Radiology, 2017, 97: 1−7. doi: 10.1016/j.ejrad.2017.10.008
[15] BODANAPALLY U K, ARCHER-ARROYO K, DREIZIN D, et al. Dual-energy computed tomography imaging of head: Virtual high-energy monochromatic (190 keV) images are more reliable than standard 120 kV images for detecting trau-matic intracranial hemorrhages[J]. Journal of Neurotrauma, 2019, 36(8): 1375−1381. doi: 10.1089/neu.2018.5985
[16] LEE S M, KIM S H, AHN S J, et al. Virtual monoenergetic dual-layer, dual-energy CT enterography: Optimization of keV settings and its added value for Crohn’s disease[J]. European Radiology, 2018, 28(6): 2525−2534. doi: 10.1007/s00330-017-5215-z
[17] SYMONS R, CHOI Y, CORK T E, et al. Optimized energy of spectral coronary CT angiography for coronary plaque detection and quantification[J]. Journal of Cardiovascular Computed Tomography, 2018, 12(2): 108−114. doi: 10.1016/j.jcct.2018.01.006
[18] APFALTRER P, SUDARSKI S, SCHNEIDER D, et al. Value of monoenergetic low-kV dual energy CT datasets for improved image quality of CT pulmonary angiography[J]. European Journal Radiology, 2014, 83(2): 322−328. doi: 10.1016/j.ejrad.2013.11.005
[19] Mc NAMARA M M, LITTLE M D, ALEXANDER L F, et al. Multireader evaluation of lesion conspicuity in small pancreatic adenocarcinomas: Complimentary value of iodine material density and low keV simulated monoenergetic images using multiphasic rapid kVp-switching dual energy CT[J]. Abdominal Imaging, 2015, 40(5): 1230−1240. doi: 10.1007/s00261-014-0274-y
[20] DARRAS K E, Mc LAUGHLIN P D, KANG H, et al. Virtual monoenergetic reconstruction of contrast-enhanced dual energy CT at 70 keV maximizes mural enhance-ment in acute small bowel obstruction[J]. European Journal Radiology, 2016, 85(5): 950−956. doi: 10.1016/j.ejrad.2016.02.019
[21] MATSUMOTO K, JINZAKI M, TANAMI Y, et al. Virtual monochromatic spectral imaging with fast kilovoltage switching: Improved image quality as compared with that obtained with conventional 120-kVp CT[J]. Radiology, 2011, 259(1): 257−62. doi: 10.1148/radiol.11100978
[22] 周旸, 曾勇明, 周蜜, 等. 双能CT单能谱70 keV图像与常规CT 120 kVp图像的等效性[J]. 中国医学影像技术, 2015,31(7): 1100−1105. ZHOU Y, CENG Y M, ZHOU M, et al. The equivalence of dual-energy CT single energy spectrum 70 keV image and conventional CT 120 kVp image[J]. China Medical Imaging Technology, 2015, 31(7): 1100−1105. (in Chinese).
-
期刊类型引用(3)
1. 刘海燕,邱晓晖,章辉庆,锁咏梅,王超,王亚丽. 能谱CT单能量成像对结直肠癌供血动脉图像质量及辐射剂量的影响. 中国CT和MRI杂志. 2025(01): 159-161 . 百度学术
2. 叶雄鑫,刘元芬,汤博荣,陈依林,郑莞怡,薛莉薇,张孝勇. 深度学习图像重建和能谱成像在低对比剂流速胸主动脉CTA中的价值. CT理论与应用研究. 2024(06): 683-691 . 本站查看
3. 何亮,唐彩银,张继,田为中,朱鹏飞. 能谱CT金属伪影抑制算法在胸部穿刺活检中的应用价值. 中国CT和MRI杂志. 2023(11): 50-52 . 百度学术
其他类型引用(1)