Intercomparison of Humidity and Temperature Sensors: GTS1, Vaisala RS80, and CFH
Authors:BIAN Jianchun  CHEN Hongbin  Holger VOMEL  DUAN Yunjun  XUAN Yuejian and LU Daren
Institution:Key Laboratory of Middle Atmosphere and Global Environment Observation, Institute of Atmospheric Physics, Chinese Academy of Sciences, Beijing 100029,Key Laboratory of Middle Atmosphere and Global Environment Observation, Institute of Atmospheric Physics, Chinese Academy of Sciences, Beijing 100029,GRUAN Lead Center, Meteorologisches Observatorium Lindenberg, German Weather Service, Germany,Kunming National Reference Climatological Station, Kunming 650034,Key Laboratory of Middle Atmosphere and Global Environment Observation, Institute of Atmospheric Physics, Chinese Academy of Sciences, Beijing 100029,Key Laboratory of Middle Atmosphere and Global Environment Observation, Institute of Atmospheric Physics, Chinese Academy of Sciences, Beijing 100029
Abstract:GTS1 digital radiosonde, developed by the Shanghai Changwang Meteorological Science and Technology Company in 1998, is now widely used in operational radiosonde stations in China. A preliminary comparison of simultaneous humidity measurements by the GTS1 radiosonde, the Vaisala RS80 radiosonde, and the Cryogenic Frostpoint Hygrometer (CFH), launched at Kunming in August 2009, reveals a large dry bias produced by the GTS1 humidity sensor. The average relative dry bias is in the order of 10% below 500 hPa, increasing rapidly to 30% above 500 hPa, and up to 55% at 310 hPa. A much larger dry bias is observed in the daytime, and this daytime effect increases with altitude. The GTS1 radiosonde fails to respond to humidity changes in the upper troposphere, and sometimes even in the middle troposphere. The failure of GTS1 in the middle and upper troposphere will result in significant artificial humidity shifts in radiosonde climate records at stations in China where a transition from mechanical to digital radiosondes has occurred. A comparison of simultaneous temperature observations by the GTS1 radiosonde and the Vaisala RS80 radiosonde suggests that these two radiosondes provide highly reproducible temperature measurements in the troposphere, but produce opposite biases for daytime and nighttime measurements in the stratosphere. In the stratosphere, the GTS1 shows a warm bias (<0.5 K) in the daytime and a relatively large cool bias (-0.2 K to -1.6 K) at nighttime.
Keywords:GTS1 radiosonde  humidity  dry bias
本文献已被 CNKI SpringerLink 等数据库收录!
点击此处可从《大气科学进展》浏览原始摘要信息
点击此处可从《大气科学进展》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号