多排螺旋CT五个剂量测量点位的统计学差异
引用本文:郑海亮,李兴东,张鹏,谈春荣,冯树理,温廷国,段永利,傅岩.多排螺旋CT五个剂量测量点位的统计学差异[J].CT理论与应用研究,2015,24(3):337-343.
作者姓名:郑海亮  李兴东  张鹏  谈春荣  冯树理  温廷国  段永利  傅岩
作者单位:1. 首都医科大学附属北京世纪坛医院医学工程处, 北京 100038;
基金项目:国家自然科学基金(11005119),中国计量科学研究院院级课题(25-JB1336-13),北京世纪坛医院院级课题(2013-C11,2013-C13).
摘    要:目的:探讨螺旋CT五个辐射剂量测量点位(CTDI100)值之间的差异和相关性统计学关系,以及该机器的辐射稳定性。方法:通过辐射剂量计和辐射剂量模体对螺旋CT五个测量点位进行12次重复测定,对各个点位的测量结果计算其相对偏差,然后用单因素方差分析(One-Way ANOVA)方法分析各点位测量结果之间的显著性。结果:5个点位重复测量的相对标准差分别为0.33%(中心点)、0.73%(3点位)、0.32%(6点位)、1.00%(9点位)和0.32%(12点位);3点位和9点位的平均值接近,单因素方差分析无显著性差异;12点位和6点位的平均值与3点位平均值的相对偏差在0.5%以下。结论:测试螺旋CT辐射稳定性良好,测试机型3点位和9点位在CTDI100上测量等效。在非验收测试和状态测试场合CTDIP可以进行简化。 

关 键 词:CT辐射剂量指数100    加权CT辐射剂量指数    测量重复性    螺旋CT    统计分析
收稿时间:2015-04-25

Statistics Difference of Five Radiation Gage Points for Multi-slice Spiral CT
Institution:1. Shijitan Hospital, Capital Medical University Medical Engineering Department, Beijing 100038, China;2. Institute of High Energy Physics, China Academy of Science, Beijing 100049, China;3. National Institute of Metrology, China, Beijing 100029, China, Beijing 100029, China;4. Shijitan Hospital, Capital Medical University Information Center, Beijing 100038, China;5. Shijitan Hospital, Capital Medical University Radiology Department, Beijing 100038, China
Abstract:Object: To investigate the statistic relationship between 5 clock positions of CTDI100 and stabilization of radiation. Method: Radiation meter and phantom are used to 12 time repeating tests on the radiation dose of a spiral CT at 5 Clock positions and then one-way ANOVA method is used to character the their difference significance of those 5 clock positions. Result: The relevant standard difference of measurements at 5 test clock positions are 0.33%(center), 0.73%(3 clock), 0.32%(6 clock), 1.00%(9 clock), 0.32%(12 clock) respectively; meanwhile the average of measurements at 3 clock is much close to that at 9 clock and no significant difference is found in one-way ANOVA test;Moreover the relative deviation between mean value of 12 Clock plus 6 Clock and 3 Clock is 0.5%. Conclusion: The tested spiral CT shows good radiation stability and 3 Clock position have equal effect with 9 clock's in CTDI100 measuring, moreover, peripheral CTDI of that can be simplified except for non-acceptance test or operational condition test. 
Keywords:CTDI100  CTDIw  repeatability  spiral CT  statistic analysis