Calibration of a chalcogenide glass membrane ion-selective electrode for the determination of free Fe in seawater: I. Measurements in UV photooxidised seawater
Authors:R De Marco  D J Mackey
Institution:a School of Applied Chemistry, Curtin University of Technology, GPO Box U 1987, Perth, Western Australia 6845, Australia;b CSIRO Marine Research, GPO Box 1538, Hobart, Tasmania 7001, Australia
Abstract:Calibration of a chalcogenide glass membrane, Fe(III)ISE Fe2.5(Ge28Sb12Se60)97.5], in buffered saline media has been undertaken in order to assess the suitability of this ISE for seawater analyses. The electrode slopes in saline citrate and salicylate buffers were 26.3 and 28.2 mV/decade, respectively, for Fe3+ concentrations ranging from 10−10 M to less than 10−25 M Fe3+. The calibration lines in the citrate and salicylate buffers were essentially collinear with the response in unbuffered chloride-free standards containing >10−5 M Fe3+, demonstrating that the response of the FeISE is unaffected by chloride ions. A mechanism involving a combination of charge transfer and ion-exchange of Fe(III), at the electrode diffusion layer, can be used to explain the ≈30 mV/decade slope of the FeISE. The response of the FeISE in UV photooxidised seawater containing 8 nM total Fe was measured as the pH was changed from 8.27 to 3.51. The slope of the response was 24.2 mV/decade Fe3+] calculated as a function of pH using Fe(III) hydrolysis constants for seawater. Moreover, the response was essentially collinear with that in citrate buffers and in unbuffered solutions containing >10−5 M Fe3+ and the slope for the combined data was 26.2 mV/decade. This study was restricted to organic-free seawater because the certainty in Fe(III)–ligand stability constants is insufficient to warrant the selection of an ideal calibration buffer system, and there is evidence that powerful chelating ligands (e.g., EDTA along with humic and fulvic acids) may alter the response of the Fe(III)ISE. The Fe dissolution rate of the FeISE in UV photooxidised seawater was found to be 1.6×10−2 nmol Fe/min, as measured by cathodic stripping voltammetry (CSV). This would contaminate a 100-ml sample by 0.8–1.6 nM Fe over a typical measurement period of 5–10 min obtained using a stability criterion of 0.5 mV/min. Various methods are proposed for reducing the level of contamination in open ocean samples that contain sub-nanomolar concentrations of iron. The FeISE has the potential to detect free Fe3+ at concentrations typically found in natural seawater.
Keywords:chalcogenide  iron  ion-selective electrode  UV
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号