首页 | 本学科首页   官方微博 | 高级检索  
     检索      

自适应统计迭代重建技术在髋关节64排CT低管电流检查中的应用
引用本文:苏大军,查云飞.自适应统计迭代重建技术在髋关节64排CT低管电流检查中的应用[J].CT理论与应用研究,2014(6):1011-1017.
作者姓名:苏大军  查云飞
作者单位:武汉大学人民医院放射科,武汉430060
摘    要:目的:探讨自适应统计迭代重建(ASIR)技术在髋关节64排CT低管电流检查中的应用。方法:106例髋关节64排螺旋CT扫描患者,随机选取53例患者采用常规管电流220 m As扫描,并采用滤过反投影(FBP)方式重建。另外53例患者采用低管电流50 m As扫描,并采用30%ASIR技术。对两组髋关节图像的辐射剂量和图像质量进行评价,并比较不同重建图像在同一层面的图像噪声值和信噪比。结果:常规管电流组与低管电流组的CT剂量指数(CTDIvol)分别为(15.55±0.20)m Gy、(3.56±0.15)m Gy,低管电流组平均CTDIvol较常规管电流组降低77.1%;常规管电流组与低管电流组的剂量长度乘积(DLP)分别为(297.82±2.9)m Gy·cm、(67.59±3.3)m Gy·cm,低管电流组平均DLP较常规管电流组降低77.3%,低管电流组与常规管电流组相比,辐射剂量大大降低(P〈0.05)。常规管电流组图像质量评分为优、良、差的患者分别为49、4和0例;低管电流组图像质量评分为优、良、差的患者分别为42、11及0例。两组图像的差异无统计学意义(P〉0.05)。图像噪声值分别为31.44、44.29,差异有统计学意义(P〈0.05)。图像信噪比分别为11.03、10.06,差异有统计学意义(P〈0.05)。结论:ASIR技术应用于64排螺旋CT髋关节低管电流扫描,大幅降低辐射剂量的同时,图像质量也能满足临床诊断的需要。

关 键 词:迭代重建算法  髋关节  辐射剂量  螺旋CT

Application of ASIR Techniques in Hip with Low Tube Current 64 Row CT
SU Da-jun,ZHA Yun-fei.Application of ASIR Techniques in Hip with Low Tube Current 64 Row CT[J].Computerized Tomography Theory and Applications,2014(6):1011-1017.
Authors:SU Da-jun  ZHA Yun-fei
Institution:(Department of Radiology, Renmin Hospital of Wuhan University, Wuhan 430060, China)
Abstract:Objective: To explore the feasibility of adaptive statistical iterative reconstruction(ASIR) technique in 64-slice CT of hip with low tube current application. Methods: Of 106 patients with 64 row helical CT scan of hip,53 patients were treated with conventional tube current of 220 m As scanning, and were reconstructed with filtered back projection(FBP), 53 patients were treated with low tube current of 50 m As scanning and were reconstructed with 30% of adaptive statistical iterative reconstruction(ASIR).The radiation dose, image noise(S D value) and signal-to-noise ratio of different reconstruction images of two groups were compared respectively at the same level. Results: The CTDIvol value was(15.55 ± 0.20) m Gy in the conventional group and(3.56 ± 0.15) m Gy in the low tube current group, the CTDIvol in the low tube current group was 77.1% of that in the conventional group, the DLP was(297.82 ± 2.9) m Gy·cm in the conventional group and(67.59 ± 3.3) m Gy·cm in the low tube current group, the DLP in the low tube current group was 77.3% of that in the conventional group. Compared with the conventional group, the radiation dose of low tube current group was reduced greatly(P 〈 0.05). The image quality with excellent、good、bad was 49, 4 and 0 cases, respectively, in the conventional group. 42, 11 and 0 cases in the low tube current group,difference of image quality between two groups were no statistically significant(P 〉 0.05). The difference of image noise value and SNR between two groups were statistically significant, respectively(P 〈 0.05). Conclusion: It is feasible to apply 30% of ASIR technology in 64 row helical CT of hip with low-current of 50 m A, resulting with the reduction of radiation dose, meanwhile, the image quality can meet the need of clinical diagnosing.
Keywords:iterative reconstruction algorithms  hip joint  radiation dose  spiral CT
本文献已被 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号