首页 | 本学科首页   官方微博 | 高级检索  
     检索      

电感耦合等离子体质谱法测定地下水中镉的不确定度评定
引用本文:许春雪,王亚平,王苏明,安子怡,代阿芳,袁建,朱家平.电感耦合等离子体质谱法测定地下水中镉的不确定度评定[J].岩矿测试,2010,29(5):607-612.
作者姓名:许春雪  王亚平  王苏明  安子怡  代阿芳  袁建  朱家平
作者单位:1. 国家地质实验测试中心,北京,100037
2. 国家地质实验测试中心,北京,100037;中国地质大学(武汉),湖北,武汉,430074
3. 中国地质调查局南京地质矿产研究所,江苏,南京,210016
基金项目:国土资源部公益性行业科研专项经费资助,国家自然科学基金项目资助,国土资源地质大调查--地下水污染测试技术研究项目资助 
摘    要:采用不确定度连续传递模型,对电感耦合等离子体质谱法测定地下水中镉元素的不确定度进行评定。其不确定度主要来源于标准溶液引入的不确定度、曲线拟合产生的不确定度和测量过程中引入的不确定度三部分。采用双误差回归的方式对标准曲线进行拟合,通过计算出各分量的不确定度,合成得到测量结果的总不确定度。

关 键 词:不确定度评定    地下水  电感耦合等离子体质谱法  双误差回归
收稿时间:5/4/2010 12:00:00 AM
修稿时间:2010/8/12 0:00:00

Uncertainty Evaluation for the Determination of Cadmium in Groundwater Samples by Inductively Coupled Plasma-Mass Spectrometry
XU Chun-xue,WANG Ya-ping,WANG Su-ming,AN Zi-yi,DAI A-fang,YUAN Jian and ZHU Jia-ping.Uncertainty Evaluation for the Determination of Cadmium in Groundwater Samples by Inductively Coupled Plasma-Mass Spectrometry[J].Rock and Mineral Analysis,2010,29(5):607-612.
Authors:XU Chun-xue  WANG Ya-ping  WANG Su-ming  AN Zi-yi  DAI A-fang  YUAN Jian and ZHU Jia-ping
Institution:National Research Center for Geoanalysis, Beijing 100037, China;National Research Center for Geoanalysis, Beijing 100037, China;National Research Center for Geoanalysis, Beijing 100037, China;National Research Center for Geoanalysis, Beijing 100037, China;National Research Center for Geoanalysis, Beijing 100037, China;China University of Geosciences, Wuhan 430074, China;National Research Center for Geoanalysis, Beijing 100037, China;Nanjing Institute of Geology and Mineral Resources, China Geological Survey, Nanjing 210016, China
Abstract:The measurement uncertainty for the analytical results of cdmium in groundwater samples by inductively coupled plasma-mass spectrometry (ICP-MS) was evaluated using Continuous Propagation Model of Uncertainty. The main sources of measurement uncertainty came from sub-uncertainties of calibration solutions, calibration curve fitting and measurements. The double-error regression was used in the course of standard curve fitting. The overall uncertainty of measurement results was obtained from the synthesis of sub-uncertainties.
Keywords:uncertainty evaluation  cadmium  groundwater  inductively coupled plasma-mass spectrometry  double-error regression
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《岩矿测试》浏览原始摘要信息
点击此处可从《岩矿测试》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏