Behaviour of quartz gravity meters under high frequency vertical motion |
| |
Authors: | Vladimír Tobyáš Karel Diviš Reviewer Z. Šimon |
| |
Affiliation: | (1) Geophysical Institute, Czechosl. Acad. Sci., Prague;(2) Geodetic and Cartographic Enterprise, Prague |
| |
Abstract: | Summary Tests on the vertical vibrating table in the frequency range of70–110 Hz indicate that quartz gravity meters are10–100 times more sensitive at some frequencies than under low-frequency excitation. At high frequencies, the reading beam is at rest and deflected from the correct position. Slow fluctuations of amplitude and frequency near resonance could cause slow irregular motion of the beam with absence of low-frequency ground motion of sufficient intensity.