RESEARCH ON DETERMINATION OF DISSOLVED OXYGEN IN SEAWATER |
| |
Authors: | ZHANG RONGKUN and LUO YINGHUA |
| |
Affiliation: | Xiamen University |
| |
Abstract: | Based on the principle of laminar flow-constant potential-chronocoulometry, this article presents a new method of determining the dissolved oxygen in seawater with considerable success. A new model of electrochemical sensor has been designed for this purpose.The sensor has the advantage of overcoming hysteresis and "outskirts effect", thus increasing its stability and prolonging its use to 1-2 years.In connection with this new method a new model SY-1 integrating microcoulorneter with digital display is designed.It is a fast (a sample/2 min) and accurate (±1-1.5%) method, suitable for use in the laboratory on board a ship. |
| |
Keywords: | |
本文献已被 CNKI 等数据库收录! |
| 点击此处可从《海洋学报(英文版)》浏览原始摘要信息 |
|
点击此处可从《海洋学报(英文版)》下载免费的PDF全文 |