Lever constants and gravity meter sensitivity |
| |
Authors: | Vladimír Tobyáš Karel Diviš Reviewer Z. Šimon |
| |
Affiliation: | (1) Geophysical Institute, Czechosl. Acad. Sci., Prague;(2) Institute of Geodesy, Prague |
| |
Abstract: | Summary For the Worden, Sharpe and Scintrex gravity meters, the lever constants were derived from the amplitude response to the vertical motion and their relation to the sensitivity of the gravity meter was verified. It is then possible to calculate the lever constants for the actual sensitivity of the gravity meter without further measurements and to apply them to estimating the dynamic behaviour of the reading beam. |
| |