排序方式: 共有1条查询结果,搜索用时 0 毫秒
1
1.
LiYan He HuaNing Qiu HeSheng Shi JunZhang Zhu XiuJuan Bai JianBing Yun 《中国科学:地球科学(英文版)》2016,59(1):111-117
Noble and active gases are released from geological samples during gas extraction for noble gas isotope analyses. The active gases should be removed before inletting to mass spectrometers for the analyses. The normal noble gas preparation systems can clean up most geological samples. However, authigenic minerals from sedimentary rocks in oil/gas fields contain organic matter, which cannot be cleaned up by the normal preparation systems and thus influence the noble gas analyses. We introduce a novel gas purification system(PRC patent No. ZL201320117751.2), which includes several reversible purification pumps with different absorbing and degassing temperatures. It can well clean up water steam, carbon dioxide and organic gases. Mica minerals are often used for ~40Ar/~39 Ar dating. A muscovite sample(2082MS) which could not be cleaned up by the normal preparation system with two SAES NP10~#174; getters, becomes the test sample for a comparative experiment in this study. The experiment is assigned into 4 sections with the organic gas removal system(OGRS) "Closed/Opened" in turn. When the OGRS is closed only with two NP10 getters for purification, the ~40Ar intensities increase in curves with inlet time because of impurities, the ~40Ar/~39 Ar dating results yield age errors about ±2%–±1%(2σ). When the OGRS is opened for purification, in contrast, the ~40 Ar intensities decrease linearly with inlet time. This indicates that the gases have been cleaned up effectively, and the 4~0Ar/~39 Ar results yield ages with errors in ±0.4%. The OGRS is very helpful to obtain high-quality analysis data. 相似文献