1.
A new technique for the in situ analysis of Re, Au, Pd, Pt and Rh in natural basalt glass by laser ablation (LA)-ICP-MS is described. The method involves external calibration against NIST SRM 612/613 or 614/615 glass certified reference materials, internal standardisation using Ca, and ablation with a 200 μm wide beam spot and a pulsed laser repetition rate of 50 Hz. Under these conditions, sensitivities for Re, Au, Pd, Pt and Rh analyte ions are ˜ 5000 to 100,000 cps/μg g-1. This is sufficient to make measurements precise to ˜ 10% at the 2-10 μg g-1 level, which is well within the range of concentrations expected in many basalts. For LA-ICP-MS calibration and a demonstration of the accuracy of the technique, concentrations of Re, Au, Pd, Pt and Rh in the NIST SRM 610/611 (˜ 1 to 50 μg g-1), 612/613 (˜ 1 to 7 μg g-1), 614/615 (˜ 0.2 to 2 μg g-1) and 616/617 (˜ 0.004 to 2 μg g-1) glasses were determined by solution-nebulisation (SN)-ICP-MS. Using the 612/613 or 614/615 glasses as calibration standards, LA-ICP-MS measurements of these elements in the other NIST glasses fell within ˜ 15% of those determined by SN-ICP-MS. Replicate LA-ICP-MS analyses of the 612/613 and 614/615 glasses indicate that, apart from certain anomalous domains, the glasses are homogeneous for Re, Au, Pd, Pt and Rh to better than 3.5%. Two LA-ICP-MS analyses of natural, island-arc basalt glasses exhibit large fractionations of Re, Au and Pd relative to Pt and Rh, compared to the relative abundances in the primitive mantle.  相似文献   

2.
A procedure for the determination of chlorine by the isotope dilution technique (ID) using negative thermal ionisation mass spectrometry (N-TIMS) is described. Silicate samples of about 10 mg were spiked and decomposed with hydrofluoric acid, and chlorine was isolated by precipitation of silver chloride after neutralisation with Ca(OH)2. The ammonical solution of AgCl was then subjected to N-TIMS. Replicate analyses of rock reference materials, typically of JB-1 and JR-1, demonstrated the high quality of the analyses (precision for Cl was ± 1-2%). We present here the most precise data sets of chlorine concentrations in nine igneous rock reference materials, three basalts (JB-1, JB-2, JB-3), two andesites (JA-3, AGV-1), two rhyolites (JR-1, JR-2) and two granodiorites (JG-3, GSP-1). The chlorine concentrations found ranged from 152 μg g-1 in AGV-1 to 1008 μg g-1 in JR-1. Our results presented here are partly (but not completely) in agreement with recommended values, where they are available. The N-TIMS ID technique can thus be used as a means of determining low chlorine contents in silicate materials to high precision.  相似文献   

3.
Trace elements in the Geological Survey of Japan carbonate reference materials Coral JCp-1 and Giant Clam JCt-1 were determined by inductively coupled plasma-mass spectrometry after digestion with 2% v/v HNO3. A standard addition method was adopted in this determination in order to neutralise the Ca matrix effect. In addition, Sc, Y, In and Bi were used as internal standards to control the matrix effect and correct instrumental drift. Of the eighteen elements measured in JCp-1, precisions for fourteen elements, including Cu, Cd and Ba, were better than 10% RSD and concentrations ranged from 0.002 μg g-1 (Cs) to 8.02 μg g-1 (Ba). The concentrations of measured trace elements in JCt-1, except for Cu, were lower than those in JCp-1. Precisions for all elements with concentrations higher than 0.04 μg g-1 in JCt-1 were also better than 10% RSD and concentrations were found to be between 0.001 μg g-1 (Cs) and 4.84 μg g-1 (Ba). The concentrations of more than fifteen trace elements in the aragonite reference materials are reported here for the first time. Both reference materials are suitable for use in geochemical studies of environmental reconstruction based upon biogenic carbonate materials.  相似文献   

4.
A combination of EMPA, sensitive high resolution ion microprobe (SHRIMP II) and/or LA-ICP-MS techniques was used to measure the concentration of selenium (Se) in NIST SRM 610, 612, 614 and a range of reference materials. Our new compiled value for the concentration of Se in NIST SRM 610 is 112 ± 2 μg g−1. The concentration of Se in NIST SRM 612, using NIST SRM 610 for calibration, determined using LA-ICP-MS (confirmed using SHRIMP II) was 15.2 ± 0.2 μg g−1. The concentration of Se in NIST SRM 614, using LA-ICP-MS was 0.394 ± 0.012 μg g−1. LA-ICP-MS determination of Se in synthetic geological glasses BCR-2G, BIR-1G, TB-1G and the MPI-DING glasses showed a range in concentrations from 0.062 to 0.168 μg g−1. Selenium in the natural glass, VG2, was 0.204 ± 0.028 μg g−1.  相似文献   

5.
We present boron isotope and concentration data from magmatic (komatiitic to rhyolitic) and sedimentary geological silicate and artificial glass reference materials that cover a wide spectrum of boron isotope compositions and boron concentrations. Boron isotope compositions were determined by TIMS (Cs2BO2+ -graphite and BO2- method) and boron concentrations by ICP-AES. Boron concentrations ranged from 7 to 159μ g-1 and agree within 14% with published values. Based on replicate analyses of individually prepared sample aliquots an overall external reproducibility of better than 10% was determined. The obtained δ11B values ranged from -12.6 to +13.6% and were reproducible within 1.1 % (2 RSD; excluding NTIMS) on the basis of individually prepared sample aliquots. The δ11B values of JA-1 (+5.3%), JB-3 (+5.9%) and JR-2 (+2.9%) overlap the published data within analytical uncertainty. For the first time δ11B values for the TB (-12.6%) and the MPI-DING glasses GOR-128-G (+13.6%), GOR-132-G (+7.1 %) and StHs6/80-G (-4.5%) are reported. The δ11B values obtained by the Cs2BO2+ -graphite and the BO2- method as well as the majority of δ11B values obtained using different sample preparation methods agree within analytical uncertainty. Therefore, we conclude that none of these analytical methods introduce any systematic error on the obtained δ11B values.  相似文献   

6.
A simple and accurate method to determine fluorine and chlorine contents in small amounts (∼ 30 mg) in rock has been developed using ion chromatography after extraction by alkaline fusion. Powdered sample was mixed with sodium carbonate and zinc oxide at a mass ratio of 1:3:1, and was fused in an electric furnace at 900 °C for 30-40 minutes. An aqueous solution obtained by dissolving the fused silicate rock was diluted to the appropriate concentration of sodium carbonate (< ∼ 24 mmol l-1) to minimise the tailing effect on F- during ion chromatography caused by the large amount of carbonate species originating from the flux. Fluorine and chlorine contents were then determined by a standard additions method. Based on the relative standard deviation of the backgrounds, detection limits of both fluorine and chlorine were ∼ 4 μg g-1, when 30 mg test portions were fused and diluted by a factor of 1200. We also report new fluorine and chlorine contents in nine GSJ (Geological Survey of Japan) reference materials, including peridotite (JP-1), granite (JG-1a), basalts (JB-1b, 2 and 3), andesites (JA-1 and 2) and rhyolites (JR-1 and 2). Fluorine and chlorine contents in the reference materials in this study were consistent with previously reported values. Reproducibilities were < 10 % for samples with F and Cl concentrations of > 20 μg g-1 and < 20 % with F and Cl < 20 μg g-1.  相似文献   

7.
A precise and simple method for the determination of lithium concentrations in small amounts of silicate sample was developed by applying isotope dilution-inductively coupled plasma-mass spectrometry (ID-ICP-MS). Samples plus a Li spike were digested with HF-HClO4, dried and diluted with HNO3, and measured by ICP-MS. No matrix effects were observed for 7Li/6Li in rock solutions with a dilution factor (DF) of 97 at an ICP power of 1.7 kW. By this method, the determination of 0.5 μg g-1 Li in a silicate sample of 1 mg can be made with a blank correction of < 1%. Lithium contents of ultrabasic to acidic silicate reference materials (JP-1, JB-2, JB-3, JA-1, JA-2, JA-3, JR-1 and JR-2 from the Geological Survey of Japan, and PCC-1 from the US Geological Survey) and chondrites (three different Allende and one Murchison sample) of 8 to 81 mg were determined. The relative standard deviation (RSD) was typically < 1.7%. Lithium contents of these samples were further determined by isotope dilution-thermal ionisation mass spectrometry (ID-TIMS). The relative differences between ID-ICP-MS and ID-TIMS were typically < 2%, indicating the high accuracy of ID-ICP-MS developed in this study.  相似文献   

8.
The USGS reference glasses GSA-1G, GSC-1G, GSD-1G, GSE-1G, BCR-2G, BHVO-2G and BIR-1G were investigated by different analytical techniques. All these materials have a geological (basaltic) matrix and are therefore useful in igneous geochemistry as matrix-matched reference materials for microanalytical techniques. The new GS glasses have trace elements in groups at concentration levels of about < 0.01, 5, 50 and 500 μg g-1. Their major element compositions have been determined by EPMA, and trace elements have been analysed by LA-ICP-MS and two isotope dilution techniques using TIMS and ICP-MS. EPMA and LA-ICP-MS analyses indicated that the USGS reference glasses are homogeneous at the μm to mm scale with respect to major (variations < 1-2%) and most trace elements (variations 1-4%). Trace element data obtained from the different analytical techniques agreed within an uncertainty of 1-5%, indicating that between method results are comparable. Therefore, the preliminary working values for the four USGS GS glasses calculated from these data have a low level of uncertainty.  相似文献   

9.
We determined chlorine contents in nine GSJ (Geological Survey of Japan) reference materials (JB-1, 1a, 2, 3; JA-1, 2, 3; JR-1, 2) by prompt gamma neutron activation analysis, employing the standard addition method. Pressed powder disks of each reference material were used for neutron irradiation and gammaray measurement, after known quantities (25-200 μl) of sodium chloride solution were added. The influence of the nearby sodium peak overlap was checked, and fluctuations in the chlorine count rate were corrected using silicon as an internal standard. The slopes of calibration lines for seven reference materials (JB-1, 2, 3; JA-1, 2, 3; JR-2) and SiO2 powders fall within 5% error, and their chlorine values were obtained from the intercepts. Chlorine contents in JB-1 a and JR-1 were also determined by using the calibration lines. Our chlorine values ranged from 26.1 to 934 μg g-1, which agrees well with the previously reported values.  相似文献   

10.
Fifty-two trace elements in NIST SRM 614, 616 and MPI-DING BM90/21-G glass reference materials as well as in NIST SRM 612, USGS BCR2-G and other MPI-DING reference glasses (KL2-G, GOR132-G, GOR128-G, ATHO-G, Tl-G, StHs6/80-G and ML3B-G) were determined by laser ablation-inductively coupled plasma-mass spectrometry (LA-ICP-MS). Accurate ultra-low trace element abundances in the NIST SRM 614, 616 and BM90/21-G reference glasses down to lower ng g−1 levels were determined with relative standard deviations (RSD) of less than 10%. Limits of detection using He as carrier gas were up to two times lower than with Ar and were 0.004 to 0.12 μg g−1 for elements of lower mass numbers (amu < 85) and 0.002 to 0.06 μg g−1 for elements having amu < 85. The measured concentrations generally agree within 15% with previous studies except for B in NIST SRM 614 and 616, which appears to be heterogeneously distributed, and Co, Zn, Ga and Ag in NIST SRM 616 for which the existing data set is too small to evaluate the discrepancies. New values for As (0.593 μg g−1), Ag (0.361 μg g−1) and Cd (0.566 μg g−1) in NIST SRM 614 and new values for Na (94864 μg g−1) and As (0.276 μg g−1) in NIST SRM 616 are reported.  相似文献   

11.
12.
We found that the suppression of signals for 88Sr, 140Ce and 238U in rock solution caused by rock matrix in ICP-MS (matrix effects) was reduced at high power operation (1.7 kW) of the ICP. To make the signal suppression by the matrix negligible, minimum dilution factors (DF) of the rock solution for Sr, Ce and U were 600, 400 and 113 at 1.1, 1.4 and 1.7 kW, respectively. Based on these findings, a rapid and precise determination method for Rb, Sr, Y, Cs, Ba, REE, Pb, Th and U using FI (flow injection)-ICP-MS was developed. The amount of the sample solution required for FI-ICP-MS was 0.2 ml, so that 1.8 mg sample was sufficient for analysis with a detection limit of several ng g-1. Using this method, we determined the trace element concentrations in the USGS rock reference materials, DTS-1, PCC-1, BCR-1 and AGV-1, and the GSJ rock reference materials, JP-1, JB-1, -2, -3, JA-1, -2 and -3. The reproducibilities (RSD %) in replicate analyses (n=5) of BCR-1, AGV-1, JB-1, -2, -3, JA-1, -2, and -3 were < 6 %, and typically 2.5%. The difference between the average concentrations of this study for BCR-1 and those of the reference values were < 2%. Therefore, it was concluded that the method can give reliable data for trace elements in silicate rocks.  相似文献   

13.
We report new data on the trace element concentrations of Mg, Cr, Mn, Co, Ni, Cu, Zn, Sr, Cd, Ba, La, Ce, Nd, Pb and U in USGS carbonate reference materials (MACS-1 and MACS-2) and compare solution ICP-MS and LA-ICP-MS trace element determinations on landfill calcites using calibration to different reference materials (MACS-1 and MACS-2 carbonate and NIST SRM 612 glass). Very good agreement (differences below 10% relative) was found between laser ablation and solution ICP-MS data for MACS-1 with higher concentrations of trace elements (values between 100 and 150 μg g−1), with the exception of Cu and Zn. Similarly good agreement was found for MACS-2 with lower trace element concentrations (units to tens of μg g−1), with the exception of Cr, Co and Zn. The MACS-1 reference material for calibration of LA-ICP-MS was found to be extremely useful for in situ determination of trace elements in real-world carbonate samples (landfill calcites), especially those present in calcite in higher concentrations (Mn, Sr, Ba; < 5% RSD). Less accurate determinations were generally obtained for trace elements present at low concentrations (∼ units of μg g−1). In addition, good agreement was observed between the instrument calibration to MACS and NIST SRM 612 glass for in situ measurements of trace elements in landfill calcites K-2, K-3 and K-4 (differences below 15% relative for most elements). Thus, the application of MACS carbonate reference materials is promising and points to the need for the development of new carbonate reference materials for laser ablation ICP-MS.  相似文献   

14.
The microanalytical capability of laser ablation microprobe-inductively coupled plasma-mass spectrometry (LAM-ICP-MS) to determine ultra trace elemental concentrations has been demonstrated by the analysis of two low concentration glass standard reference materials, NIST SRM 614 and 616. Results for fifty two elements at concentrations in the low ng g-1 range are compared with those determined using secondary ion mass spectrometry (SIMS). Both techniques provide results at these concentrations that generally agree within 95% confidence limits, demonstrating the accuracy for ultra-trace level of in situ determinations by the two techniques. At concentrations of less than 20 ng g-1 in NIST SRM 616, an accuracy and precision of better than 10% has been obtained for most mono-isotopic rare earth elements, when a spot size of 50 μm is used. Limits of detection for selected elements were as low as 0.5 ng g-1.  相似文献   

15.
The analytical capabilities of laser ablation (LA)-ICP-MS in determining Li, Be and B at trace levels in geological samples have been tested on a series of glass reference materials and natural samples. The LA-ICP-MS instrument used consisted of a sector-field ICP-MS coupled with a laser ablation microprobe operating at either 266 or 213 nm wavelength. Reference glasses from NIST (SRM 612, 614 and 616) and MPI-DING (KL2-G, ML3B-G, StHs6/80-G, GOR128-G, GOR132-G, T1-G and ATHO-G) were selected to develop the analytical method and to assess the best instrumental configuration. A series of calcic amphiboles with different Li, Be and B concentrations were also analysed using both LA-ICP-MS and SIMS to test the applicability of the method to natural minerals. Results indicated that with a spot size of 40 μm the agreement between measured and reference values of Li, Be and B is generally better than 10% for NIST SRM 612 and 20% for NIST SRM 614. Average reproducibility at the 2s level was 10% for Li, 20% for Be and 15% for B. Limits of detection were approximately 100 ng g-1 for Be and B and 200 ng g-1 for Li. These results were confirmed by analyses carried out on natural amphiboles and compared well in terms of precision and accuracy with those commonly achieved by SIMS.  相似文献   

16.
Three new certified reference materials (CRM), certified for the platinum-group elements (PGE), GPt-8, GPt-9 and GPt-10 were developed based on the previous CRMs IGGE GPt-1 to GPt-7. The PGE concentration of GPt-8 is about 1 ng g-1. GPt-9 and GPt-10 are ore samples with PGE concentrations of more than 1 μg g-1. A multi-laboratory collaborative analysis scheme was adopted in the certification procedure, in which nine highly-experienced institutes and laboratories participated. The samples were analysed for the six platinum-group elements by nickel sulfide mini fire assay, with Te coprecipitation, and were determined by ICP-MS. Osmium was determined by isotope dilution.  相似文献   

17.
Procedures for sampling, sample preparation and ICP-MS analysis of endemic sponges from Lake Baikal have been developed. Sample decomposition was carried out using an open acid decomposition with ultrasound treatment. The distribution of nineteen elements (Mg, Al, P, Ca, Ti, Mn, Co, Ni, Cu, Rb, Sr, Y, Cd, Ba, La, Ce, Pb, Th and U) in different parts of a sponge's body (outer and inner layers and layers adjacent to the substratum) was studied. Detection limits were determined; these ranged from 0.013 to 4.12 μg g-1 for trace elements and from 23 to 130 μg g-1 for biogenic elements. The degree of elemental uptake by living substances is discussed with regard to the environment.  相似文献   

18.
The platinum-group elements (PGE) and gold have been determined in twenty international rock reference materials by inductively coupled plasma-mass spectrometry (ICP-MS) after pre-concentration by a nickel sulfide fire assay. It was possible to achieve determination limits for a 50 g sample that ranged from 1 pg g-1 (Rh) to 23 pg g-1 (Au). Compared to published certified and recommended values for rock reference materials, the trueness of the method was found to be good. However, in some cases we observed large deviations for all elements in the sub 10 ng g-1 range within individual reference sample splits. Our results show that the PGE and Au are inhomogeneously distributed in the reference materials analysed here, where they are present in low concentrations, using 50 g test portions.  相似文献   

19.
The paper presents preliminary results of the use of a high resolution double-focussing, magnetic sector inductively coupled plasma-mass spectrometer (HR-ICP-MS) with ultraviolet laser ablation (LA) for the bulk analysis of geological materials fused with Li2B4O7. Detection limits are based on data from precision measurements of a fused SiO2 sample of high purity, and sensitivity data (cps/μg g-1) obtained on the Reference Material (RM) Syenite SY-2. For many trace elements, the detection limits are better than 0.05 μg g-1 using a sample to flux weight ratio of 1:7.
Calibration curves, which are based entirely on RMs, are established for Hf, Ta, Tb, Tm and Lu. They indicate that, even at this early stage in the development of the technique, data accurate to ˜ 25% can be collected. It is concluded that the method may prove to be a valuable supplement to XRF for low level element concentration measurements; it is also very practical, as the same sample discs can be used for both XRF and LA-ICP-MS analyses.  相似文献   

20.
The beryllium and zirconium contents of 45 geochemical reference samples have been determined by inductively coupled plasma after fusion of the samples with lithium metaborate and dissolution of the melt in dilute nitric acid. The method described here is rapid and sample preparation straightforward. Good agreement is shown with previously published results for these two elements. A correction has to be made for an interference due to vanadium in determining the beryllium content, and there is a slight interference due to yttrium in the determination of zirconium. The detection limit for beryllium is about 0.2 μg g-1 and for zirconium about 15 μg g-1 in the sample.  相似文献   