Quick Search:       Advanced Search
WANG Na,XU Tie-min,WEI Shuang,WANG Jia-song,ZENG Jiang-ping,ZHANG Nan.Determination of Rare Earth Elements in Ultra-fine Rock and Soil Samples by ICP-MS Using Microwave Digestion[J].Rock and Mineral Analysis,2020,39(1):68-76
View Full Text   View/Add Comment   EndNote   BibTeX   Download reader
Determination of Rare Earth Elements in Ultra-fine Rock and Soil Samples by ICP-MS Using Microwave Digestion
Received:April 01, 2019  Revised:May 28, 2019
DOI:10.15898/j.cnki.11-2131/td.201904010043
Key words: rock  soil  rare earth elements  ultra-fine sample preparation  wet-milling  ethanol  microwave digestion  inductively coupled plasma-mass spectrometry
Author NameAffiliation
WANG Na Tianjin Center of Geological Survey, China Geological Survey, Tianjin 300170, China
Key Laboratory of Muddy Coast Geo-Environment, China Geological Survey, Tianjin 300170, China 
XU Tie-min Tianjin Center of Geological Survey, China Geological Survey, Tianjin 300170, China 
WEI Shuang Tianjin Center of Geological Survey, China Geological Survey, Tianjin 300170, China 
WANG Jia-song Tianjin Center of Geological Survey, China Geological Survey, Tianjin 300170, China 
ZENG Jiang-ping Tianjin Center of Geological Survey, China Geological Survey, Tianjin 300170, China 
ZHANG Nan Tianjin Center of Geological Survey, China Geological Survey, Tianjin 300170, China 
Hits: 210
Download times: 135
Abstract:
      BACKGROUND: The sampling weight can be as low as milligram level when the contents of rare earth elements in rock and soil samples are determined by inductively coupled plasma-mass spectrometry (ICP-MS). However, the size of 200 mesh samples requires 100mg to ensure representativeness, which makes it difficult for ICP-MS to make full use of its advantages of high sensitivity and small sampling volume.
OBJECTIVES: To establish the method for the determination of rare earth elements in ultra-fine rock and soil samples by ICP-MS with microwave digestion.
METHODS: A method was developed for the determination of rare earth elements in ultra-fine rock and soil samples by ICP-MS with microwave digestion. The ultra-fine rock and soil samples were prepared by the planetary superfine pulverizing machine with wet grinding. Some parameters influencing the preparation of ultra-fine rock and soil samples, including grinding method, grinding time and the proportion of grinding balls were optimized. Microwave digestion of ultrafine-grained rock and soil samples in a nitric acid-hydrofluoric acid-hydrogen peroxide solution system, the sampling amount was reduced to 5mg, the total acid amount was reduced to 0.25mL and the digestion time was reduced to 25 minutes.
RESULTS: The results indicated that the samples were best refined when the wet-milling was adopted with ethanol as a dispersant and the milling time was 10 minutes and 15 minutes for 200 mesh soil and rock samples, respectively. The material ratios were 7g materials, 500 grinding balls, 45mL dispersants for the preparation of ultra-fine soil samples, whereas 5g materials, 500 grinding balls, 45mL dispersants for the preparation of ultra-fine rock samples. The particle size distribution D95 of soil standard substance GBW07404 and GBW07447 can be as low as 7.51μm and 7.05μm, and the D95 of rock standard substance GBW07104 and GBW07121 can be 8.42μm and 8.30μm under the optimized conditions. Microwave digestion of ultrafine-grained rock sample GBW07104 in a nitric acid-hydrofluoric acid-hydrogen peroxide solution system, the sampling amount was reduced to 5mg, the total acid amount was reduced to 0.25mL, the digestion time was reduced to 25 minutes, and the measured value of rare earth elements was consistent with the certified value with the relative standard deviation is between 1.64% and 5.21%.
CONCLUSIONS: This method was used for the detection of rare earth elements in other ultra-fine standard materials (GBW07404, GBW07447 and GBW07121), which yielded the relative standard deviation of 0.17%-6.60%, meeting the first criterion of Geology and Minerals Laboratory Testing Quality Management Standards.